(b)(4).Submission of this report does not constitute an admission that the importer or manufacturer's product caused or contributed to the event.Device evaluation by manufacturer: a field service engineer (fse) visited the customer to address the reported event.During the evaluation, fse confirmed the problem.To resolve the issue, fse adjusted the sys-in1 parameter from 128 to 119.Fse then ran a full rack of whole blood patient samples and all total area values were less than 2000.Fse ran quality controls (qc) with acceptable results.The instrument was operating as expected.There was no further action required by fse.A 13-month complaint history review and service history review for similar complaints was performed for serial number (b)(4) from 13-aug-2017 through aware date 13-sept2018.There were no similar complaints identified during the searched period.The g8 variant analysis mode operator's manual under chapter 1, introduction and applications, states the following: 1.8 limitations of the procedure- total area: dilution studies demonstrate that the assay is linear from a total area of 500 to 4000.However, the optimum total area is 700 to 3000.Interpretation of results: results will not be reported if the total area (ta) is <500 which can be seen in severe anemia.Results will not be reported if the ta is >4000 which can be seen in polycythemia.(see "abnormal red cell survival in previous section).The optimal goal for total area is between 700-3000.However a ta in the range of 500-4000 is acceptable and reportable for whole blood specimens.The most probable cause of the reported event was due to misalignment fo the sys-in1 parameter setting.
|