BAXTER HEALTHCARE - MEDOLLA MODENA MONITORS PHOENIX; SET, TUBING, BLOOD, WITH AND WITHOUT ANTI-REGURGITATION VALVE
|
Back to Search Results |
|
Catalog Number 101323 |
Device Problem
Smoking (1585)
|
Patient Problem
No Patient Involvement (2645)
|
Event Date 09/15/2018 |
Event Type
malfunction
|
Manufacturer Narrative
|
Should additional relevant information become available, a supplemental report will be submitted.
|
|
Event Description
|
It was reported that, there was a smoke and a strong burnt smell of plastic coming out from a phoenix machine while the machine was in standby mode.There was no alarm triggered when the event occurred.There was no patient involvement.No additional information is available.
|
|
Manufacturer Narrative
|
Additional information was added.The actual device was not available; however, a photograph of the sample was provided for evaluation.A visual inspection of the photographic samples was performed and it was noted that the electro valve ev1 was burnt in the self-extinguish plastic part.The reported condition was verified.The cause of the reported condition was component failure.A service history review revealed no indication that the parts replaced during servicing caused or contributed to the reported event.Should additional relevant information become available, a supplemental report will be submitted.
|
|