• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: APOLLO ENDOSURGERY, INC. LAP-BAND SYSTEM; ADJUSTABLE GASTRIC BAND

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

APOLLO ENDOSURGERY, INC. LAP-BAND SYSTEM; ADJUSTABLE GASTRIC BAND Back to Search Results
Device Problem Device Slipped (1584)
Patient Problems Adhesion(s) (1695); Ischemia (1942); Necrosis (1971); Internal Organ Perforation (1987); Pain (1994); Renal Failure (2041); Vomiting (2144); Hernia (2240); Peritonitis (2252)
Event Date 01/01/2017
Event Type  Injury  
Manufacturer Narrative
Unknown taper.The reporter of the event was asked if the product was available for analysis and to provide device information.To date, apollo has not been able to confirm device information from the reporter and the device manufacturer is unknown.Further information has been requested of the reporter regarding: implant date, explant date, date of occurrence, and additional patient information.To date, no additional information has been received by apollo.A review of the device labeling notes the following: precautions: it is the responsibility of the surgeon to advise the patient of the dietary restrictions that follow this procedure and to provide diet and behavior modification support.Failure to adhere to the dietary restrictions may result in obstruction and/or failure to lose weight.Over-dissection of the stomach during placement may result in slippage or erosion of the band and require reoperation.Do not push the tip of any instrument against the stomach wall or use excessive electrocautery.Stomach perforation or damage may result in peritonitis and death.Adverse events: it is important to discuss all possible complications and adverse events with your patient.Complications which may result from the use of this product include the risks associated with the medications and methods utilized in the surgical procedure, the risks associated with any surgical procedure and the patient's degree of intolerance to any foreign object implanted in the body.Perforation of the stomach can occur.Death can also occur.Specific complications of laparoscopic surgery can include spleen damage (sometimes requiring splenectomy) or liver damage, bleeding from major blood vessels, lung problems, thrombosis, and rupture of the wound slippage and/or pouch dilatation of the band can occur.Gastroesophageal reflux, nausea and/or vomiting with early or minor slippage may be in some cases successfully resolved by band deflation.More serious slippages may require band repositioning and/or removal.If there is total stoma outlet obstruction that does not respond to band deflation, or if there is abdominal pain, then immediate re-operation to remove the band is indicated.Nausea and vomiting may occur, particularly in the first few days after surgery and when the patient eats more than recommended.Nausea and vomiting may also be symptoms of stoma obstruction or a band/ stomach slippage.Frequent, severe vomiting can result in pouch dilatation, stomach slippage or esophageal dilatation.Deflation of the band is immediately indicated in all of these situations.Deflation of the band may alleviate excessively rapid weight loss and nausea and vomiting.Reoperation to reposition or remove the device may be required.Warnings: patients should be advised that the lap-band ap system is a long-term implant.Explant and replacement surgery may be indicated at any time.Medical management of adverse reactions may include explantation.Revision surgery for explantation and replacement may also be indicated to achieve patient satisfaction.
 
Event Description
Literature review performed: "laparoscopic management of gastric perforation secondary to mesenteroaxial volvulus in a patient with laparoscopic adjustable gastric banding." paya-liorente c, et al.Asian j endosc surg.2018 mar 7.Case presentation: patient "was admitted to the emergency department with a 2-week history of alimentary vomiting, which was occurring with increasing frequency." the patient "had undergone laparoscopic adjustable gastric banding 6 years earlier." an initial examination revealed epigastric pain.Laboratory findings showed acute renal failure and an abdominal ct scan showed band slippage, leading to the formation of mesenteroaxial gastric volvulus and ischemia on the gastric wall."emergent diagnostic laparoscopy was performed, and severe peritonitis and gastric necrosis caused by volvulation was found." it was observed that there was a herniation of the greater curvature through the band and subsequent volvulus."the rotation's axis had firm adhesion between the banding's pseudocapsule and the gastrohepatic ligament." "after band removal, a fundal perforation was noted." fourteen months later the patient remained asymptomatic.Although the manufacturer of the device is unknown, it is apollo's approach to compliance to resolve all doubt in favor of reporting.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LAP-BAND SYSTEM
Type of Device
ADJUSTABLE GASTRIC BAND
Manufacturer (Section D)
APOLLO ENDOSURGERY, INC.
1120 s. capital of texas hwy
bldg 1, ste. 300
austin TX 78746
Manufacturer (Section G)
ALLERGAN
global park free zone
900 global park
la aurora de heredia,
CS  
Manufacturer Contact
laura leboeuf
1120 s. capital of texas hwy
bldg 1, ste. 300
austin, TX 78746
MDR Report Key7953929
MDR Text Key123362691
Report Number3006722112-2018-00258
Device Sequence Number1
Product Code LTI
Combination Product (y/n)N
Reporter Country CodeSP
PMA/PMN Number
P000008
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type foreign,health professional,l
Reporter Occupation Physician
Type of Report Initial
Report Date 09/10/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 09/10/2018
Initial Date FDA Received10/10/2018
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
Patient Age54 YR
-
-