MAUDE Adverse Event Report: BOSTON SCIENTIFIC CORPORATION PROMUS PREMIER SELECT; STENT, CORONARY, DRUG-ELUTING

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC CORPORATION PROMUS PREMIER SELECT; STENT, CORONARY, DRUG-ELUTING Back to Search Results
Model Number 10665
Device Problems Difficult to Remove (1528); Device Contamination with Chemical or Other Material (2944)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 09/03/2018
Event Type  malfunction  
Manufacturer Narrative
Device is a combination product.
 
Event Description
It was reported that foreign matter was found.The 90% stenosed target lesion was located in the highly calcified ight anterior interventricular artery.Pre-dilation was performed with a 2.50mm x 20mm emerge balloon catheter and a 24 x 2.75 promus premier select stent was successfully implanted.During the withdrawal of the stent delivery system there was resistance felt at the end of the non-bsc guide catheter.When the stent delivery system was pulled off the straight tip 190cm samurai guidewire, resistance was felt again and then a thread approximately 15cm long came out and was rolled up on the side and was described as looking like plastic.The procedure was completed and there was no harm to the patient.
 
Event Description
It was reported that foreign matter was found.The 90% stenosed target lesion was located in the highly calcified ight anterior interventricular artery.Pre-dilation was performed with a 2.50mm x 20mm emerge balloon catheter and a 24 x 2.75 promus premier select stent was successfully implanted.During the withdrawal of the stent delivery system there was resistance felt at the end of the non-bsc guide catheter.When the stent delivery system was pulled off the straight tip 190cm samurai guidewire, resistance was felt again and then a thread approximately 15cm long came out and was rolled up on the side and was described as looking like plastic.The procedure was completed and there was no harm to the patient.The promus premier select stent delivery system was returned for analysis with a piece of foreign material (fm) wrapped in cloth returned separate to the device.The stent was not returned for analysis as it was implanted in the patient.A visual and microscopic examination of the bumper tip showed signs of damage.The balloon cones were reviewed and no issues were noted.The balloon folds appear relaxed from their original folded position and it appears that positive pressure was applied and a vacuum pulled.A visual and tactile examination of the hypotube found multiple hypotube kinks along the full length of the catheter.A visual and tactile examination of the shaft polymer extrusion found no issues along the midshaft, inner or outer shaft polymer extrusion.The fm was returned separate to the device wrapped in a piece of cloth.The fm measured 245 mm in length.The fm was analyzed using fourier transform infra-red spectroscopy.The spectrum produced would suggest that the fm is polytetrafluoroethylene (ptfe).
 
Manufacturer Narrative
Device is a combination product.
 
Manufacturer Narrative
Device is a combination product.Root cause was changed from 67 no problem detected to 4311 adverse event related to procedure.This supplemental correction report is being filed to correct the report number in a previously submitted supplemental report (follow-up number 001) which was accepted by fda on 01/15/2019 08:45 am ct.The report number is being corrected from: 2134265-2018-64265to: 2134265-2018-61515.
 
Event Description
It was reported that foreign matter was found.The 90% stenosed target lesion was located in the highly calcified ight anterior interventricular artery.Pre-dilation was performed with a 2.50mm x 20mm emerge balloon catheter and a 24 x 2.75 promus premier select stent was successfully implanted.During the withdrawal of the stent delivery system there was resistance felt at the end of the non-bsc guide catheter.When the stent delivery system was pulled off the straight tip 190cm samurai guidewire, resistance was felt again and then a thread approximately 15cm long came out and was rolled up on the side and was described as looking like plastic.The procedure was completed and there was no harm to the patient.The promus premier select stent delivery system was returned for analysis with a piece of foreign material (fm) wrapped in cloth returned separate to the device.The stent was not returned for analysis as it was implanted in the patient.A visual and microscopic examination of the bumper tip showed signs of damage.The balloon cones were reviewed and no issues were noted.The balloon folds appear relaxed from their original folded position and it appears that positive pressure was applied and a vacuum pulled.A visual and tactile examination of the hypotube found multiple hypotube kinks along the full length of the catheter.A visual and tactile examination of the shaft polymer extrusion found no issues along the midshaft, inner or outer shaft polymer extrusion.The fm was returned separate to the device wrapped in a piece of cloth.The fm measured 245 mm in length.The fm was analyzed using fourier transform infra-red spectroscopy.The spectrum produced would suggest that the fm is polytetrafluoroethylene (ptfe).
 
Event Description
It was reported that foreign matter was found.The 90% stenosed target lesion was located in the highly calcified right anterior interventricular artery.Pre-dilation was performed with a 2.50mm x 20mm emerge balloon catheter and a 24 x 2.75 promus premier select stent was successfully implanted.During the withdrawal of the stent delivery system there was resistance felt at the end of the non-bsc guide catheter.When the stent delivery system was pulled off the straight tip 190cm samurai guidewire, resistance was felt again and then a thread approximately 15cm long came out and was rolled up on the side and was described as looking like plastic.The procedure was completed and there was no harm to the patient.The promus premier select stent delivery system was returned for analysis with a piece of foreign material (fm) wrapped in cloth returned separate to the device.The stent was not returned for analysis as it was implanted in the patient.A visual and microscopic examination of the bumper tip showed signs of damage.The balloon cones were reviewed and no issues were noted.The balloon folds appear relaxed from their original folded position and it appears that positive pressure was applied and a vacuum pulled.A visual and tactile examination of the hypotube found multiple hypotube kinks along the full length of the catheter.A visual and tactile examination of the shaft polymer extrusion found no issues along the midshaft, inner or outer shaft polymer extrusion.The fm was returned separate to the device wrapped in a piece of cloth.The fm measured 245 mm in length.The fm was analyzed using fourier transform infra-red spectroscopy.The spectrum produced would suggest that the fm is polytetrafluoroethylene (ptfe).
 
Manufacturer Narrative
Device is a combination product.Root cause was changed from cause not established to no problem detected.This supplemental correction report is being filed to correct the report number in a previously submitted supplemental report (follow-up number 001) which was accepted by fda on 12/14/2018 12:45 pm ct.The report number is being corrected from: 2134265-2018-64265 to: 2134265-2018-61515.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PROMUS PREMIER SELECT
Type of Device
STENT, CORONARY, DRUG-ELUTING
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
two scimed place
maple grove MN 55311
MDR Report Key7956372
MDR Text Key123405134
Report Number2134265-2018-61515
Device Sequence Number1
Product Code NIQ
Combination Product (y/n)N
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Type of Report Initial,Followup,Followup,Followup
Report Date 01/25/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received10/11/2018
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date04/26/2020
Device Model Number10665
Device Catalogue Number10665
Device Lot Number0022094008
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer09/26/2018
Date Manufacturer Received12/17/2018
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Treatment
ABBOTT VISTA BRIGHT GUIDE CATHETER; ABBOTT VISTA BRIGHT GUIDE CATHETER; ABBOTT VISTA BRIGHT GUIDE CATHETER; ABBOTT VISTA BRIGHT GUIDE CATHETER; ABBOTT VISTA BRIGHT GUIDE CATHETER
Patient Age52 YR
-
-