• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: TERUMO CARDIOVASCULAR SYSTEMS CORPORATION STERILE FX25RWC W/ RES; BLOOD GAS OXYGENATOR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

TERUMO CARDIOVASCULAR SYSTEMS CORPORATION STERILE FX25RWC W/ RES; BLOOD GAS OXYGENATOR Back to Search Results
Model Number 3CX*FX25RWC
Device Problem Coagulation in Device or Device Ingredient (1096)
Patient Problem Death (1802)
Event Date 09/20/2018
Event Type  Death  
Manufacturer Narrative
Terumo has not received the device for evaluation; therefore, the investigation has yet to be completed.Terumo plans on submitting a follow-up report when the investigation is complete and when more information becomes available.For this reason, terumo references evaluation conclusion code.(b)(4).Results: results pending completion of investigation.Conclusions: conclusion not yet available.
 
Event Description
The user facility reported to terumo cardiovascular that during cardiopulmonary bypass, a presence of white and red clot was seen in the arterial filter of the oxygenator.The procedure was a aortic valve replacement and cab x 1 to right coronary artery.Heparinization: 40,000 u loading dose, 10,000 u in pump prime.Pre-bypass act was 550 seconds.Serial acts: 1st on bypass, 764 secs, then in succession with no times noted, 616, 634 and 493 seconds.An additional 10,000 u heparin given in response to the 493 seconds.The resultant act was 666 seconds.The total pump time was 1 hour and 49 minutes, with the cross clamp time being 1 hour and 34 minutes.The lowest temperature on bypass was 34.4 degrees c.The patient received tranexamic acid per routine.Prior to the termination of cardiopulmonary bypass the patient¿s hematocrit was increased from 31% to 36% by the use of a hemoconcentration.1000ml of effluent was removed from the patient¿s circulating blood volume.With the embolic concerns regarding the finding of clot formation in the oxygenator, an initial ct scan was performed early during the icu stay that showed potential changes in the hypothalamus.The patient was unconscious and hemiplegic, a follow up ct scan was performed and revealed multiple emboli in the brain.This scan was performed 3 days post operative.The lipid profile, coagulation studies performed post operative showed no abnormalities.Upon entry to the icu, the troponin level was 205, and elevated to 900 the next morning.The patient passed away 3 days post operative.Product was changed out.
 
Manufacturer Narrative
This follow-up report is submitted to fda in accord with applicable regulations ¿ and as indicated by terumo cardiovascular systems in the initial report submitted to the fda on (b)(6) 2018.Upon further investigation of the reported event, the following information is new and/or changed: (describe event or problem - corrected information on changed out); (device availability - added date returned to manufacturer); (date application evaluated by manufacturer;) (indication that this is a follow-up report); (follow-up due to correction and additional information) ; (device evaluation anticipated by manufacturer - a second follow-up will be submitted upon completion of the investigation and/or submission of new information, thus tcvs).All available information has been placed on file in quality management for appropriate tracking, trending, and follow-up.
 
Event Description
The product was not changed out.
 
Manufacturer Narrative
Upon further investigation of the reported event, the following information is new and/or changed: (additional device information - added exp date); (date received by manufacturer); (indication that this is a follow-up report); (follow-up due to additional information and device evaluation); (device evaluated by manufacturer); (device manufacture date); (identification of evaluation codes (b)(4)).The actual sample was returned for evaluation, and some clots were noted to have formed on the surface of the sample.It was visually inspected upon receipt and found that the purge line had been cut off.No other anomalies noted on the device.The actual sample was rinsed and dried.It was subsequently built into a circuit with tubes.Bovine blood was circulated in the circuit, while the pressure drop was determined at each flow rate.The obtained values were confirmed to meet the factory's specification, and no obstruction was confirmed.After the pressure test, bovine blood was kept circulated in the circuit for 6 hours with no obstruction.When the circulation was ceased, it was flushed with water, and no clot was found inside the sample.Review of the device history record and the product release control sheet of the involved product/lot number combination confirmed that, there were no indication of production-related anomalies.The evaluation of the actual sample after having been rinsed was the normal product presenting no issue relating to the confirmation of clots; therefore, the cause of this complaint cannot be determined definitely.All available information has been placed on file in quality management for appropriate tracking, trending, and follow-up.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
STERILE FX25RWC W/ RES
Type of Device
BLOOD GAS OXYGENATOR
Manufacturer (Section D)
TERUMO CARDIOVASCULAR SYSTEMS CORPORATION
125 blue ball road
elkton MD 21921
Manufacturer (Section G)
SAME
Manufacturer Contact
cathleen hargreaves
125 blue ball road
elkton, MD 21921
8002837866
MDR Report Key7957413
MDR Text Key123481847
Report Number1124841-2018-00259
Device Sequence Number1
Product Code DTZ
UDI-Device Identifier00699753450837
UDI-Public(01)00699753450837
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K151791
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,distri
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup,Followup
Report Date 12/04/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received10/11/2018
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date06/30/2021
Device Model Number3CX*FX25RWC
Device Catalogue NumberN/A
Device Lot NumberWH30
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/17/2018
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Date Manufacturer Received11/28/2018
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured07/30/2018
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Death;
Patient Age67 YR
Patient Weight100
-
-