(b)(4).One (1) unit of subassembly ph12153 033 neb adaptor phantom holder was received for analysis.This subassembly is part of the 031-33j nebulizer adaptor 033, sterile, (b)(6), related to this customer complaint.Signs of use were observed, and it was observed that adaptor doesn't spin freely.The sample was functionally tested and during the setup of the oxygen entrainment test it was observed that the assembly of the nut adaptor and the upper body was unstable.Even with that condition, the sample was able to be tested but the testing failed.After the testing finished, the adaptor was carefully disassembled from the upper body and it was visually inspected.During the visual inspection it was found there was wear on the internal tabs.Based on the investigation performed the complaint is confirmed.Although the condition was observed on the sample provided, there is not sufficient evidence to assure that this issue was originated during the manufacturing assembly or molding process.The wear on the internal tabs of the adaptor was most likely caused by the end user during the connection of the adaptor into the flowmeter that causes an unstable connection issue.
|