Analysis was performed on the returned generator.No performance or any other type of adverse conditions were found with the pulse generator.Analysis was performed on the returned lead.During the visual analysis several abraded openings were observed on the outer silicone tubing.The abraded openings found on the outer silicone tubing, most likely provided the leakage path for the dried remnants of what appeared to have once been body fluids inside the outer silicone tubing.Abrasions were also found on the connector pin inner silicone tubing.With the exception of the abraded openings on the outer and connector pin inner silicone tubing, the condition of the returned lead portion is consistent with those that typically exist following an explant procedure.No other obvious anomalies were noted.Note that since the electrode array section was not returned, an evaluation and resulting commentary cannot be made on that portion of the lead no additional relevant information has been received to date.
|