MAKO SURGICAL CORP. FEMORAL TRIAL STD LM/RL SIZE 5; PROSTHESIS, KNEE PATELLOFEMOROTIBIAL, PARTIAL, SEMI-CONSTRAINED, CEMENTED, POLYM Back to Search Results
Catalog Number 170505
Device Problem Material Deformation (2976)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 09/13/2018
Event Type  malfunction  
Manufacturer Narrative
It was noted that the device is not available for evaluation.Should additional information become available, it will be provided in a supplemental report upon completion of the investigation.Device not returned.
 
Event Description
Screw hole stripped on femoral trial.Case type: pka.
 
Manufacturer Narrative
Reported event: an event regarding crack/fracture involving a mako trial was reported.The event was confirmed.Method & results: device evaluation and results: visual inspection and material analysis: an inspection of the returned device by a materials analyst noted "damage observed on threads consistent with crossthreading and off axis loading.Based on the given information, no identifiable materials or manufacturing discrepancies were observed on the surfaces examined." medical records received and evaluation: n records were made available for evaluation.Device history review: device history review indicated the devices accepted into final stock from the reported lot were free from discrepancies.Complaint history review: there have been no other similar reported events for the reported lot.Conclusions: no further investigation is required at this time.The root cause determined because insufficient information was provided.Further information such as operative notes are needed to investigate this event further.If additional information becomes available, this investigation will be reopened.
 
Event Description
Screw hole stripped on femoral trial.Case type: pka.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
FEMORAL TRIAL STD LM/RL SIZE 5
Type of Device
PROSTHESIS, KNEE PATELLOFEMOROTIBIAL, PARTIAL, SEMI-CONSTRAINED, CEMENTED, POLYM
Manufacturer (Section D)
MAKO SURGICAL CORP.
2555 davie road
fort lauderdale FL 33317
MDR Report Key7958383
MDR Text Key123505642
Report Number3005985723-2018-00580
Device Sequence Number1
Product Code NPJ
UDI-Device Identifier00848486003500
UDI-Public00848486003500
Combination Product (y/n)N
PMA/PMN Number
K090763
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative
Type of Report Initial,Followup
Report Date 12/19/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number170505
Device Lot Number26010911
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/14/2018
Initial Date Manufacturer Received 09/13/2018
Initial Date FDA Received10/11/2018
Supplement Dates Manufacturer Received11/22/2018
Supplement Dates FDA Received12/19/2018
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient Outcome(s) Other;
-
-