• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ERIKA DE REYNOSA, S.A. DE C.V. CUSTOM COMBI SET; SET, TUBING, BLOOD, WITH AND WITHOUT ANTI-REGURGITATION VALVE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ERIKA DE REYNOSA, S.A. DE C.V. CUSTOM COMBI SET; SET, TUBING, BLOOD, WITH AND WITHOUT ANTI-REGURGITATION VALVE Back to Search Results
Catalog Number 03-2722-9
Device Problem Fluid/Blood Leak (1250)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 09/21/2018
Event Type  malfunction  
Manufacturer Narrative
The plant investigation is in process.A supplemental mdr will be submitted upon completion of this activity. .
 
Event Description
A user facility clinic manager reported a bloodline separation that occurred at the beginning of the patient¿s hemodialysis (hd) treatment.The separation was observed at the connection between the heparin line (which was leaking) and the arterial line.There was no damage noticed on the bloodlines.The patient¿s estimated blood loss (ebl) was approximately 5 ml, as most of their blood was returned.There was no patient injury, adverse events, or medical intervention required as a result of this event.The patient successfully completed treatment on the same machine with new supplies.The bloodlines are available for return to the manufacturer., harm, or adverse event was reported at intake.
 
Manufacturer Narrative
Plant investigation: the sample was not returned to the manufacturer and the lot number was not provided.A manufacturing review was performed on the products shipped to the patient for the three (3) month time frame which immediately preceded the event occurrence date.This review included the lot numbers for all fresenius combiset bloodlines shipped to this account within the selected time frame.The entire set of lots have been sold and distributed.There were no non-conformances or abnormalities identified during the manufacturing process which could be associated with the reported event.An investigation of the device history records (dhr) was conducted and confirmed that the results of the in-progress and final quality control (qc) testing met all requirements.The product lots involved met all specifications for release.A review of the dhr did not reveal a probable cause for the customer complaint.As a physical evaluation could not be performed, a definitive conclusion regarding the reported incident could not be reached and a cause could not be confirmed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CUSTOM COMBI SET
Type of Device
SET, TUBING, BLOOD, WITH AND WITHOUT ANTI-REGURGITATION VALVE
Manufacturer (Section D)
ERIKA DE REYNOSA, S.A. DE C.V.
mike allen #1331
parque industrial reynosa
reynosa 88780
MX  88780
Manufacturer (Section G)
ERIKA DE REYNOSA, S.A. DE C.V.
mike allen #1331
parque industrial reynosa
reynosa 88780
MX   88780
Manufacturer Contact
thomas c. johnson
920 winter st.
waltham, MA 02451
7816999499
MDR Report Key7964715
MDR Text Key123681142
Report Number8030665-2018-01589
Device Sequence Number1
Product Code FJK
UDI-Device Identifier00840861100293
UDI-Public00840861100293
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K962081
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Nurse
Type of Report Initial,Followup
Report Date 10/22/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number03-2722-9
Device Lot Number18ER01133
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Device Age MO
Initial Date Manufacturer Received 09/27/2018
Initial Date FDA Received10/15/2018
Supplement Dates Manufacturer Received10/19/2018
Supplement Dates FDA Received10/22/2018
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-