Model Number 397002-001
Device Problem Mechanical Problem (1384)
Patient Problem No Patient Involvement (2645)
Event Date 09/26/2018
Event Type  malfunction  
Manufacturer Narrative
This alleged failure mode poses a low risk to a patient because the issue was observed when the companion 2 driver was not supporting at patient.The companion 2 driver has been returned to syncardia for evaluation.The results of the evaluation will be provided in a follow-up mdr.Ce (b)(4) initial.
 
Event Description
The companion 2 driver was not supporting a patient.The customer, a syncardia certified hospital, reported that the key was found bent in the key switch when the companion 2 driver was delivered.The key broke off inside the key switch when hospital personnel tried to turn on the driver.
 
Manufacturer Narrative
Visual inspection of the driver revealed a broken key inside the key switch, confirming the customer-reported issue.Review of the driver's patient file data revealed no alarms indicative of a device malfunction or related to the customer-reported issue.Although the key was broken inside the key switch, the switch was still operational.The root cause of the broken key could not be conclusively determined but was likely caused by shipping conditions of the driver.Syncardia has an open corrective and preventive action (capa) to address this issue of broken keys.Syncardia has completed its evaluation and is closing this file.Ce 4484 follow-up report 1.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SYNCARDIA COMPANION 2 DRIVER
Type of Device
EXTERNAL PNEUMATIC DRIVER
Manufacturer (Section D)
SYNCARDIA SYSTEMS, LLC
1992 e. silverlake road
tucson AZ 85713
MDR Report Key7965910
MDR Text Key123806282
Report Number3003761017-2018-00480
Device Sequence Number1
Product Code LOZ
UDI-Device Identifier00858000003107
UDI-Public(01)00858000003107
Combination Product (y/n)N
PMA/PMN Number
P030011
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional,user faci
Type of Report Initial,Followup
Report Date 10/25/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received10/15/2018
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number397002-001
Device Catalogue Number397002-001
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/01/2018
Date Manufacturer Received09/26/2018
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-