Model Number 397002-001 |
Device Problem
Mechanical Problem (1384)
|
Patient Problem
No Patient Involvement (2645)
|
Event Date 09/26/2018 |
Event Type
malfunction
|
Manufacturer Narrative
|
This alleged failure mode poses a low risk to a patient because the issue was observed when the companion 2 driver was not supporting at patient.The companion 2 driver has been returned to syncardia for evaluation.The results of the evaluation will be provided in a follow-up mdr.Ce (b)(4) initial.
|
|
Event Description
|
The companion 2 driver was not supporting a patient.The customer, a syncardia certified hospital, reported that the key was found bent in the key switch when the companion 2 driver was delivered.The key broke off inside the key switch when hospital personnel tried to turn on the driver.
|
|
Manufacturer Narrative
|
Visual inspection of the driver revealed a broken key inside the key switch, confirming the customer-reported issue.Review of the driver's patient file data revealed no alarms indicative of a device malfunction or related to the customer-reported issue.Although the key was broken inside the key switch, the switch was still operational.The root cause of the broken key could not be conclusively determined but was likely caused by shipping conditions of the driver.Syncardia has an open corrective and preventive action (capa) to address this issue of broken keys.Syncardia has completed its evaluation and is closing this file.Ce 4484 follow-up report 1.
|
|