• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BOSTON SCIENTIFIC CORPORATION CAROTID WALLSTENT; STENT, CAROTID

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC CORPORATION CAROTID WALLSTENT; STENT, CAROTID Back to Search Results
Model Number 26606
Device Problems Premature Activation (1484); Retraction Problem (1536)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 08/28/2018
Event Type  malfunction  
Event Description
It was reported that the device was unable to be re-constrained initially.The target lesion was located in the severely tortuous internal carotid artery (ica).A 6.0-22mm carotid wallstent was advanced for treatment; however, difficulty in manipulating the outer sheath was encountered during deployment and the system got dislodged due to severe angulation in the common iliac artery and severe tortuosity in ica.It was tried to resheath, but severe resistance was felt, and after the system was pulled back to the common carotid artery, it was able to perform the resheathing.The device was replaced with another of the same device and the procedure was completed.No patient complications were reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CAROTID WALLSTENT
Type of Device
STENT, CAROTID
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
two scimed place
maple grove MN 55311
Manufacturer (Section G)
BOSTON SCIENTIFIC IRELAND LIMITED
ballybrit business park
galway
EI  
Manufacturer Contact
sonali arangil
two scimed place
maple grove, MN 55311
6515827403
MDR Report Key7966561
MDR Text Key123804791
Report Number2134265-2018-61844
Device Sequence Number1
Product Code NIM
Combination Product (y/n)N
Reporter Country CodeJA
PMA/PMN Number
P050019
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Physician
Type of Report Initial
Report Date 10/15/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date09/18/2021
Device Model Number26606
Device Catalogue Number26606
Device Lot Number0021883887
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 10/01/2018
Initial Date FDA Received10/15/2018
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured03/20/2018
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-