It was reported stent placement difficulties occurred.A stenting procedure was being performed in the superficial femoral artery (sfa) ostium using a retrograde popliteal approach.As the innova 6 x 20 x 130 stent was positioned and deployment initiated, the stent moved forward and partially covered the profunda artery.It was noted the physician thought the device was being held steady in position and was surprised that the stent moved forward.The stent was post dilated and the procedure was completed with no adverse effect to the patient outcome.It was noted there was no need to do anything further, the physician just didn't like the final position of the stent.Device evaluated by manufacturer: returned product consisted of an innova self-expanding stent delivery system (sds).A visual examination identified a kink at the nosecone.The stent was not returned.Per the customers statement the stent was deployed in the patient.Inspection of the remainder of the device, apart from the observed damage, revealed no other damage or irregularities.
|