• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: EXACTECH, INC. EQUINOXE REVERSE HUMERAL ADAPTER TRAY

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

EXACTECH, INC. EQUINOXE REVERSE HUMERAL ADAPTER TRAY Back to Search Results
Model Number EQUINOXE REVERSE TRAY ADAPTER PLATE TRAY +0
Device Problems Loose or Intermittent Connection (1371); Device Dislodged or Dislocated (2923)
Patient Problem Pain (1994)
Event Date 10/11/2018
Event Type  Injury  
Manufacturer Narrative
Pending evaluation.
 
Event Description
Index surgery: (b)(6) 2018.Revision due to dislocation.Patient showed up to dr.With a painful reverse shoulder.X-rays revealed a dislocated reverse, which could not be a closed reduced so an open reduction was booked.During surgery it was decided to remove liner and tray, but the stem loosened when removing humeral adaptor tray.A trial stem was inserted and a new tray/liner trailed.Once surgeon was happy the definitive implants were put in place, shoulder was reduced, and surgeon expects patient to recover well.
 
Manufacturer Narrative
Engineering evaluation noted that the revision reported was likely the result of loosening supporting ligaments and muscles, which allowed for dislocation.However, this cannot be confirmed as the explants were not available for evaluation.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EQUINOXE REVERSE HUMERAL ADAPTER TRAY
Type of Device
ADAPTER TRAY
Manufacturer (Section D)
EXACTECH, INC.
2320 nw 66th court
gainesville FL 32653
Manufacturer (Section G)
EXACTECH, INC.
2320 nw 66th court
gainesville FL 32653
Manufacturer Contact
graham cuthbert
2320 nw 66th court
gainesville, FL 32653
MDR Report Key7979436
MDR Text Key124271532
Report Number1038671-2018-00812
Device Sequence Number1
Product Code KWT
Combination Product (y/n)N
Reporter Country CodeUS
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 03/01/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Model NumberEQUINOXE REVERSE TRAY ADAPTER PLATE TRAY +0
Device Catalogue Number320-10-00
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Initial Date Manufacturer Received 10/16/2018
Initial Date FDA Received10/18/2018
Supplement Dates Manufacturer Received10/16/2018
Supplement Dates FDA Received03/01/2019
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? Yes
Type of Device Usage Initial
Patient Sequence Number1
Patient Age54 YR
-
-