BOSTON SCIENTIFIC CORPORATION DYNAGEN CRT-D; DEFIBRILLATOR, AUTOMATIC IMPLANTABLE CARDIOVERTER, WITH CRT-D
|
Back to Search Results |
|
Model Number G151 |
Device Problems
High impedance (1291); Incorrect, Inadequate or Imprecise Result or Readings (1535); Failure to Convert Rhythm (1540); Device Displays Incorrect Message (2591)
|
Patient Problems
No Known Impact Or Consequence To Patient (2692); No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 07/31/2018 |
Event Type
Injury
|
Manufacturer Narrative
|
As no further information concerning this report is expected, our investigation is complete.This investigation will be updated should further information be provided.
|
|
Event Description
|
Boston scientific received information that, during a device change out procedure, the right ventricular (rv) lead was connected to this new cardiac resynchronization therapy defibrillator (crt-d) and an alert for high out of range shock impedances was observed.The values were greater than 200 ohms.The physician then performed a commanded shock and a code 1005 indicative of a shock lead open condition triggered.The physician elected to continue to monitor the patient as the patient's ejection fraction has increased and with their previous device, they had not received any therapy.The procedure concluded and the system remains in service.No additional adverse patient effects were reported.
|
|
Event Description
|
Boston scientific received information that, during a device change out procedure, the right ventricular (rv) lead was connected to this new cardiac resynchronization therapy defibrillator (crt-d) and an alert for high out of range shock impedances was observed.The values were greater than 200 ohms.The physician then performed a commanded shock and a code 1005 indicative of a shock lead open condition triggered.The physician elected to continue to monitor the patient as the patient's ejection fraction has increased and with their previous device, they had not received any therapy.The procedure concluded and the system remains in service.No additional adverse patient effects were reported.It was reported that the patient was seen and noninvasive programmed stimulation (nips) was performed.Daily measurements showed shock impedance measurements to be 107 ohms in triad.The patient was noted to have been lost to follow up for some time.An arrhythmia was successfully induced during testing, however, the device was unsuccessful in converting the rhythm and the patient was externally rescued.Surgical intervention was undertaken one week later.This cardiac resynchronization therapy defibrillator (crt-d) was explanted and replaced and the right ventricular (rv) defibrillation lead was surgically abandoned and replaced.No additional adverse patient effects were reported.This product has not been returned for analysis.This report will be updated should additional information become available or if analysis is completed.
|
|
Manufacturer Narrative
|
The returned device was thoroughly inspected and analyzed upon receipt at our post market quality assurance laboratory.Visual examination of the device header and case noted no anomalies.Pin gage testing, designed to verify proper port dimensions, was completed.Each port measured as expected.The device was then exposed to simulated heart load conditions, and the defibrillation, pacing, and sensing functions were tested.Impedance testing was completed and all measurements were within normal limits.The device operated appropriately with no interruptions in therapy output at the returned programmed settings.A series of electrical tests was also performed, and again, normal device function was observed.Analysis did not identify any device characteristics that would have caused or contributed to the reported clinical observations.
|
|
Event Description
|
Boston scientific received information that, during a device change out procedure, the right ventricular (rv) lead was connected to this new cardiac resynchronization therapy defibrillator (crt-d) and an alert for high out of range shock impedances was observed.The values were greater than 200 ohms.The physician then performed a commanded shock and a code 1005 indicative of a shock lead open condition triggered.The physician elected to continue to monitor the patient as the patient's ejection fraction has increased and with their previous device, they had not received any therapy.The procedure concluded and the system remains in service.No additional adverse patient effects were reported.It was reported that the patient was seen and noninvasive programmed stimulation (nips) was performed.Daily measurements showed shock impedance measurements to be 107 ohms in triad.The patient was noted to have been lost to follow up for some time.An arrhythmia was successfully induced during testing, however, the device was unsuccessful in converting the rhythm and the patient was externally rescued.Surgical intervention was undertaken one week later.This cardiac resynchronization therapy defibrillator (crt-d) was explanted and replaced and the right ventricular (rv) defibrillation lead was surgically abandoned and replaced.No additional adverse patient effects were reported.This product has not been returned for analysis.This report will be updated should additional information become available or if analysis is completed.
|
|
Search Alerts/Recalls
|
|
|