• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ZIMMER BIOMET, INC. G7 NEUTRAL E1 LINER 32MM F; PROSTHESIS, HIP

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ZIMMER BIOMET, INC. G7 NEUTRAL E1 LINER 32MM F; PROSTHESIS, HIP Back to Search Results
Model Number N/A
Device Problem Device Markings/Labelling Problem (2911)
Patient Problem No Information (3190)
Event Date 07/27/2018
Event Type  Injury  
Manufacturer Narrative
(b)(4).(b)(6).Customer has indicated that the product is in process of being returned to zimmer biomet for investigation.Once the investigation has been completed, a follow-up mdr will be submitted.
 
Event Description
It was reported that while opening the implant during operation, incorrect size was noted.There was a delay of 50 minutes due to having to open a new product.No additional information was available.
 
Event Description
No further event information available at the time of this report.
 
Manufacturer Narrative
This follow-up report is being submitted to relay additional information.Complaint sample was evaluated and the reported event was not confirmed.Device was returned and evaluated.Dimensional analysis of key features outside diameter and over all cup height verified product is correct size.In addition, the label content attached to mhr was verified, and found label to be correct for item, lot, size and description.Dhr was reviewed and no discrepancies were found.Root cause was unable to be determined.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
G7 NEUTRAL E1 LINER 32MM F
Type of Device
PROSTHESIS, HIP
Manufacturer (Section D)
ZIMMER BIOMET, INC.
56 e. bell drive
warsaw IN 46582
MDR Report Key7981420
MDR Text Key124289411
Report Number0001825034-2018-09824
Device Sequence Number1
Product Code PBI
Combination Product (y/n)N
PMA/PMN Number
PK121874
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,consum
Type of Report Initial,Followup
Report Date 07/01/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date05/12/2021
Device Model NumberN/A
Device Catalogue Number010000850
Device Lot Number3794196
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/19/2018
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 09/20/2018
Initial Date FDA Received10/18/2018
Supplement Dates Manufacturer Received06/28/2019
Supplement Dates FDA Received07/02/2019
Is This a Reprocessed and Reused Single-Use Device? No
Removal/Correction NumberN/A
Patient Sequence Number1
Patient Outcome(s) Hospitalization; Other;
-
-