• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: GOMCO DEVICE FOR CIRCUMCISION 1.1 CLAMP; CLAMP, CIRCUMCISION

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

GOMCO DEVICE FOR CIRCUMCISION 1.1 CLAMP; CLAMP, CIRCUMCISION Back to Search Results
Device Problem Sharp Edges (4013)
Patient Problem Hemorrhage/Bleeding (1888)
Event Date 07/25/2018
Event Type  malfunction  
Event Description
Physician was doing a circumcision using the gomco device with a 1.1 clamp.As the procedure was completed they noted bleeding from the side of the baby's penis.Pressure was applied, followed by silver nitrate and eventually stitches were required to stop the bleeding.Upon close inspection of the gomco device, it was noted that the "bell" had a sharp edge.We don't have a method to track how old the device might have been or how many times it has been used and reprocessed.As a result, two gomco devices with 1.1 clamps were taken out of service to avoid this from happening again.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
GOMCO DEVICE FOR CIRCUMCISION 1.1 CLAMP
Type of Device
CLAMP, CIRCUMCISION
MDR Report Key7981789
MDR Text Key124555379
Report Number7981789
Device Sequence Number1
Product Code HFX
Combination Product (y/n)N
Number of Events Reported1
Summary Report (Y/N)N
Report Source User Facility
Reporter Occupation Risk Manager
Type of Report Initial
Report Date 10/02/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? No
Was the Report Sent to FDA? Yes
Date Report Sent to FDA10/02/2018
Event Location Hospital
Date Report to Manufacturer10/18/2018
Initial Date Manufacturer Received Not provided
Initial Date FDA Received10/18/2018
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage N
Patient Sequence Number1
Patient Outcome(s) Other;
Patient Age2 DA
-
-