BOSTON SCIENTIFIC CORPORATION EASYTRAK 2 IS-1; DEFIBRILLATOR, AUTOMATIC IMPLANTABLE CARDIOVERTER, WITH CRT-D
|
Back to Search Results |
|
Model Number 4542 |
Device Problems
Human-Device Interface Problem (2949); Adverse Event Without Identified Device or Use Problem (2993)
|
Patient Problem
Vascular Dissection (3160)
|
Event Date 08/27/2018 |
Event Type
Injury
|
Manufacturer Narrative
|
At this time, the product has not been returned.If the product is returned, analysis will be performed and this report would be updated at that time.
|
|
Event Description
|
Boston scientific received information that this left ventricular (lv) lead was not successfully implanted after it dissected the posterior lateral vein.Since there were no other veins available to place the lead, the lead was removed and the patient received an epicardial lead.
|
|
Manufacturer Narrative
|
Upon receipt at our post market quality assurance laboratory, a thorough evaluation of the lead was performed.Resistance and pressure tests were completed to assess lead electrical performance and insulation integrity.Measurements throughout these tests were within normal limits.Microscopic inspections of the terminal pin assembly, lead body, and electrode tip found no anomalies.Laboratory testing was unable to reproduce the reported clinical observations and detailed analysis did not reveal any abnormalities.Laboratory analysis did not identify any lead characteristics that would have caused or contributed to the reported clinical observations.
|
|