Model Number G148
Device Problems High impedance (1291); Over-Sensing (1438); Ambient Noise Problem (2877)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 08/29/2018
Event Type  malfunction  
Manufacturer Narrative
The device remains in service at this time.If additional information is received, this report will be updated.
 
Event Description
Boston scientific received information that this cardiac resynchronization therapy defibrillator (crt-d) and a non-boston scientific right atrial (ra) lead exhibited noise and oversensing of minute ventilation (mv) signals.The ra impedance measurements fluctuated between 500 and 2000 ohms.The mv feature was programmed off.No adverse patient effects were reported.The device remains in service.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
INOGEN
Type of Device
IMPLANTABLE CHF GENERATOR
Manufacturer (Section D)
GUIDANT CRM CLONMEL IRELAND
guidant ireland
clonmel, tipperary ireland
Manufacturer (Section G)
GUIDANT CRM CLONMEL IRELAND
guidant ireland
clonmel, tipperary ireland
Manufacturer Contact
tim degroot
4100 hamline ave. n
st. paul, MN 
6515826168
MDR Report Key7984124
MDR Text Key124358371
Report Number2124215-2018-18533
Device Sequence Number1
Product Code NIK
UDI-Device Identifier00802526534591
UDI-Public(01)00802526534591(17)20170911
Combination Product (y/n)N
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 08/29/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Expiration Date09/11/2017
Device Model NumberG148
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? Yes
Initial Date Manufacturer Received 08/29/2018
Initial Date FDA Received10/19/2018
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured09/11/2015
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Treatment
1882TC; G148; MISMATCH
Patient Age61 YR
-
-