Boston scientific received information that the patient with this cardiac resynchronization therapy defibrillator (crt-d) received inappropriate atrial tachy response (atr) due to noise and oversensing on the right atrial (ra) lead.In addition, the ra lead pacing impedances had intermittently spiked, however, they were still within normal limits.The patient experienced palpitations.The patient's ra lead was a non-boston scientific product.Boston scientific technical services (ts) discussed that the noise was respiratory rate trend (rrt) oversensing and recommended programming it off.They also mentioned there was questionable ra capture and recommended evaluating the lead.The electrogram (egm) indicated that the patient didn't experience greater than 2 seconds of asystole.The minute ventilation (mv) sensor was programmed off and the noise was immediately removed.This crt-d system remains in service.No adverse patient effects were reported.
|