• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SYNCARDIA SYSTEMS, LLC SYNCARDIA FREEDOM HOSPITAL AC POWER SUPPLY

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SYNCARDIA SYSTEMS, LLC SYNCARDIA FREEDOM HOSPITAL AC POWER SUPPLY Back to Search Results
Catalog Number 295400-002
Device Problem Electrical /Electronic Property Problem (1198)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 10/05/2018
Event Type  malfunction  
Manufacturer Narrative
This alleged failure mode poses a low risk to the patient because the reported issue did not prevent the freedom driver from performing its life-sustaining functions.In addition, the freedom driver has a redundant power source of multiple onboard batteries.The freedom hospital ac power supply will be returned to syncardia for evaluation.The results of the investigation will be provided in a follow-up mdr.(b)(4).
 
Event Description
The freedom hospital ac power supply is a component that enables the freedom driver to be plugged into an external power source.The customer, a syncardia authorized distributor, reported that the clinic informed them that the freedom hospital ac power supply would not charge onboard batteries inside the freedom driver when plugged into a wall.The customer also reported that the patient was provided with a replacement ac power supply.
 
Manufacturer Narrative
The ac power supply did not pass functional testing as it produced an inappropriately low output voltage per specification and was not able to power a driver for a minimum of one minute, thus confirming the customer-reported issue.The root cause cannot be determined without physical access to the internal printed circuit board.This issue will continue to be monitored and trended as part of the customer experience process.Syncardia has completed its evaluation and is closing this file.Ce 4499 follow-up report 1.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SYNCARDIA FREEDOM HOSPITAL AC POWER SUPPLY
Type of Device
AC POWER SUPPLY
Manufacturer (Section D)
SYNCARDIA SYSTEMS, LLC
1992 e. silverlake road
tucson AZ 85713
MDR Report Key7995416
MDR Text Key124814684
Report Number3003761017-2018-00491
Device Sequence Number1
Product Code LOZ
Combination Product (y/n)N
PMA/PMN Number
P030011
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type distributor,foreign,other
Type of Report Initial,Followup
Report Date 01/28/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received10/23/2018
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number295400-002
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/07/2019
Date Manufacturer Received10/05/2018
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient Age46 YR
-
-