Catalog Number 295400-002 |
Device Problem
Electrical /Electronic Property Problem (1198)
|
Patient Problem
No Consequences Or Impact To Patient (2199)
|
Event Date 10/05/2018 |
Event Type
malfunction
|
Manufacturer Narrative
|
This alleged failure mode poses a low risk to the patient because the reported issue did not prevent the freedom driver from performing its life-sustaining functions.In addition, the freedom driver has a redundant power source of multiple onboard batteries.The freedom hospital ac power supply will be returned to syncardia for evaluation.The results of the investigation will be provided in a follow-up mdr.(b)(4).
|
|
Event Description
|
The freedom hospital ac power supply is a component that enables the freedom driver to be plugged into an external power source.The customer, a syncardia authorized distributor, reported that the clinic informed them that the freedom hospital ac power supply would not charge onboard batteries inside the freedom driver when plugged into a wall.The customer also reported that the patient was provided with a replacement ac power supply.
|
|
Manufacturer Narrative
|
The ac power supply did not pass functional testing as it produced an inappropriately low output voltage per specification and was not able to power a driver for a minimum of one minute, thus confirming the customer-reported issue.The root cause cannot be determined without physical access to the internal printed circuit board.This issue will continue to be monitored and trended as part of the customer experience process.Syncardia has completed its evaluation and is closing this file.Ce 4499 follow-up report 1.
|
|