Catalog Number 295053-001 |
Device Problem
Electrical /Electronic Property Problem (1198)
|
Patient Problem
No Consequences Or Impact To Patient (2199)
|
Event Date 10/08/2018 |
Event Type
malfunction
|
Manufacturer Narrative
|
This alleged failure mode poses a low risk to the patient because it does not prevent the freedom driver from performing its life-sustaining functions.The freedom driver has a redundant power source of onboard batteries.The freedom car charger will be returned to syncardia for evaluation.The results of the evaluation will be provided in a follow-up mdr.(b)(4).
|
|
Event Description
|
The freedom car charger is component that enables the freedom driver to be plugged into a 12v vehicle power outlet.The customer, a syncardia certified hospital, reported that the patient's freedom car charger did not work.
|
|
Manufacturer Narrative
|
The customer-reported issue of the car charger not working was confirmed via the functional inspection, where the led did not illuminate and no voltage passed through the unit.Internal inspection of the car charger revealed a blown fuse.When the fuse was replaced with a fuse from a known operational freedom car charger, functionality was restored, including illumination of the status indicator led and passing 13.985vdc to the output of the device.This issue will continue to be monitored and trended as part of the customer experience process.Syncardia has completed its evaluation and is closing this file.(b)(4) follow-up report 1.
|
|