Model Number 595000-001 |
Device Problem
Image Display Error/Artifact (1304)
|
Patient Problem
No Patient Involvement (2645)
|
Event Date 10/02/2018 |
Event Type
malfunction
|
Manufacturer Narrative
|
This alleged failure mode poses a low risk to a patient because the issue was observed when the freedom driver was not supporting a patient.The freedom driver has been returned to syncardia for evaluation.The results of the investigation will be provided in a follow-up mdr.(b)(4).
|
|
Event Description
|
The freedom driver was not supporting a patient.The customer reported that freedom driver display showed asterisks and not numbers for the fill volume and cardiac output.
|
|
Manufacturer Narrative
|
The driver in "as received" condition passed all test requirements associated with normotensive and hypertensive settings.The customer-reported issue (driver showing asterisks and not numbers for the fill volume and cardiac output) has historically been produced when the plugs in the driveline connectors are not removed from the cpc connectors before powering on the driver.Test plugs were inserted into the connectors and the driver was powered on.This testing reproduced the customer-reported issue.The driver performed as intended with no evidence of a device malfunction.This issue will continue to be monitored and trended as part of the customer experience process.Syncardia has completed its evaluation and is closing this file.(b)(4) follow-up report 1.
|
|