Model Number 595000-001 |
Device Problem
Audible Prompt/Feedback Problem (4020)
|
Patient Problem
No Patient Involvement (2645)
|
Event Date 10/02/2018 |
Event Type
malfunction
|
Manufacturer Narrative
|
This alleged failure mode poses a low risk to the patient because the issue was observed while the freedom driver was not supporting a patient.The freedom driver has been returned to syncardia for evaluation.The results of the investigation will be provided in a follow-up mdr.(b)(4).
|
|
Event Description
|
The freedom driver was not supporting a patient.The customer, a syncardia certified hospital, reported that the freedom driver instantly exhibited a fault alarm when turned on, one day after passing system check out.
|
|
Manufacturer Narrative
|
The customer-reported issue of a fault alarm upon start-up was confirmed via review of the driver's alarm history and testing.The alarm history was reviewed and revealed alarm codes produced as a result of the primary motor not engaging for greater than or equal to 5 seconds after continuous open or short detected.During investigation testing, the driver started up on the secondary motor and enunciated an alarm as designed.The root cause of the customer-reported issue was determined to be a malfunction of the main printed circuit board assembly (pcba) not powering the primary motor piston cylinder assembly (pca) triggering the driver to convert to secondary motor operation.Syncardia has a corrective and preventive action (capa) for this issue.Syncardia has completed its evaluation and is closing this file.Ce 4503 follow-up report 1.
|
|