• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SYNCARDIA SYSTEMS, LLC SYNCARDIA FREEDOM DRIVER; EXTERNAL PNEUMATIC DRIVER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SYNCARDIA SYSTEMS, LLC SYNCARDIA FREEDOM DRIVER; EXTERNAL PNEUMATIC DRIVER Back to Search Results
Model Number 595000-001
Device Problem Audible Prompt/Feedback Problem (4020)
Patient Problem No Patient Involvement (2645)
Event Date 10/02/2018
Event Type  malfunction  
Manufacturer Narrative
This alleged failure mode poses a low risk to the patient because the issue was observed while the freedom driver was not supporting a patient.The freedom driver has been returned to syncardia for evaluation.The results of the investigation will be provided in a follow-up mdr.(b)(4).
 
Event Description
The freedom driver was not supporting a patient.The customer, a syncardia certified hospital, reported that the freedom driver instantly exhibited a fault alarm when turned on, one day after passing system check out.
 
Manufacturer Narrative
The customer-reported issue of a fault alarm upon start-up was confirmed via review of the driver's alarm history and testing.The alarm history was reviewed and revealed alarm codes produced as a result of the primary motor not engaging for greater than or equal to 5 seconds after continuous open or short detected.During investigation testing, the driver started up on the secondary motor and enunciated an alarm as designed.The root cause of the customer-reported issue was determined to be a malfunction of the main printed circuit board assembly (pcba) not powering the primary motor piston cylinder assembly (pca) triggering the driver to convert to secondary motor operation.Syncardia has a corrective and preventive action (capa) for this issue.Syncardia has completed its evaluation and is closing this file.Ce 4503 follow-up report 1.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SYNCARDIA FREEDOM DRIVER
Type of Device
EXTERNAL PNEUMATIC DRIVER
Manufacturer (Section D)
SYNCARDIA SYSTEMS, LLC
1992 e. silverlake road
tucson AZ 85713
MDR Report Key7995423
MDR Text Key124815115
Report Number3003761017-2018-00494
Device Sequence Number1
Product Code LOZ
UDI-Device Identifier00858000003121
UDI-Public(01)00858000003121
Combination Product (y/n)N
PMA/PMN Number
P030011
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional,user faci
Type of Report Initial,Followup
Report Date 01/04/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number595000-001
Device Catalogue Number595000-001
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/09/2018
Initial Date Manufacturer Received 10/03/2018
Initial Date FDA Received10/23/2018
Supplement Dates Manufacturer Received10/03/2018
Supplement Dates FDA Received01/07/2019
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-