Model Number N/A |
Device Problem
Difficult to Insert (1316)
|
Patient Problem
No Known Impact Or Consequence To Patient (2692)
|
Event Date 09/26/2018 |
Event Type
Injury
|
Manufacturer Narrative
|
(b)(4).Product has been received by zimmer biomet and the investigation is in process.Once the investigation has been completed, a follow-up mdr will be submitted.
|
|
Event Description
|
It was reported that during a right total hip arthroplasty, the liner would not lock into the shell.Another liner that was the same part number and size was fitted with no issues.Attempts have been made and additional information on the reported event is unavailable at this time.
|
|
Manufacturer Narrative
|
This follow-up report is being submitted to relay additional information.(b)(4).Complaint sample was evaluated and the reported event was not confirmed.Visual inspection of the returned liner showed damage on the scallop due to impaction.No other anomaly was observed.Automated measurements taken at the time of production determined that the device was conforming to print specifications.Device history record (dhr) was reviewed and no discrepancies relevant to the reported event were found.Root cause was unable to be determined as the necessary information to adequately investigate the reported event was not provided.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
|
|
Event Description
|
No additional event information to report at this time.
|
|