• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ZIMMER BIOMET, INC. G7 HI-WALL ARCOMXL LNR 40MM G; PROSTHESIS, HIP

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ZIMMER BIOMET, INC. G7 HI-WALL ARCOMXL LNR 40MM G; PROSTHESIS, HIP Back to Search Results
Model Number N/A
Device Problem Difficult to Insert (1316)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 09/26/2018
Event Type  Injury  
Manufacturer Narrative
(b)(4).Product has been received by zimmer biomet and the investigation is in process.Once the investigation has been completed, a follow-up mdr will be submitted.
 
Event Description
It was reported that during a right total hip arthroplasty, the liner would not lock into the shell.Another liner that was the same part number and size was fitted with no issues.Attempts have been made and additional information on the reported event is unavailable at this time.
 
Manufacturer Narrative
This follow-up report is being submitted to relay additional information.(b)(4).Complaint sample was evaluated and the reported event was not confirmed.Visual inspection of the returned liner showed damage on the scallop due to impaction.No other anomaly was observed.Automated measurements taken at the time of production determined that the device was conforming to print specifications.Device history record (dhr) was reviewed and no discrepancies relevant to the reported event were found.Root cause was unable to be determined as the necessary information to adequately investigate the reported event was not provided.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Event Description
No additional event information to report at this time.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
G7 HI-WALL ARCOMXL LNR 40MM G
Type of Device
PROSTHESIS, HIP
Manufacturer (Section D)
ZIMMER BIOMET, INC.
56 e. bell drive
warsaw IN 46582
MDR Report Key7996710
MDR Text Key124830242
Report Number0001825034-2018-09865
Device Sequence Number1
Product Code PBI
Combination Product (y/n)N
PMA/PMN Number
K121874
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type distributor,health profession
Type of Report Initial,Followup
Report Date 11/21/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date01/04/2023
Device Model NumberN/A
Device Catalogue Number010000826
Device Lot Number6197083
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/11/2018
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 09/27/2018
Initial Date FDA Received10/23/2018
Supplement Dates Manufacturer Received11/15/2018
Supplement Dates FDA Received11/21/2018
Is This a Reprocessed and Reused Single-Use Device? No
Removal/Correction NumberN/A
Patient Sequence Number1
-
-