• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BECTON DICKINSON MEDICAL SYSTEMS BD LUER-LOK¿ SYRINGE STERILE, SINGLE USE; PISTON SYRINGE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BECTON DICKINSON MEDICAL SYSTEMS BD LUER-LOK¿ SYRINGE STERILE, SINGLE USE; PISTON SYRINGE Back to Search Results
Catalog Number 309646
Device Problem Difficult to Remove (1528)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 10/03/2018
Event Type  malfunction  
Manufacturer Narrative
A device evaluation is anticipated, but has not yet begun.Upon completion of the investigation, a supplemental report will be filed.
 
Event Description
It was reported that a bd luer-lok¿ syringe sterile, single use had a plunger that was hard to pull back.There was no report of exposure, serious injury, or medical intervention.
 
Manufacturer Narrative
Investigation summary: since no samples displaying the condition reported are available for examination, we were unable to fully investigate this incident, therefore a root cause could not be determined.Dhr review for batch 8035612 ((b)(4).Manufacturing dates: 02/12/2018 to 02/14/2018.Batch quantity was 480,000.All visual inspections were performed as per requirement with no quality notifications related to the complaint defect.Batch 8035612 was inspected and accepted based on meeting our inspection control plan and subsequently approved for shipment.The review showed no rejected inspections or quality issues during the production of the provided lot number that could have contributed to the reported defect.Complaints received for this device and reported condition will continue to be tracked and trended.Information will be captured on trend reports and monitored monthly.Our business team regularly reviews the collected data for identification of emerging trends.
 
Event Description
It was reported that a bd luer-lok¿ syringe sterile, single use had a plunger that was hard to pull back.There was no report of exposure, serious injury, or medical intervention.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BD LUER-LOK¿ SYRINGE STERILE, SINGLE USE
Type of Device
PISTON SYRINGE
Manufacturer (Section D)
BECTON DICKINSON MEDICAL SYSTEMS
route 7 and grace way
canaan CT 06018
MDR Report Key7996885
MDR Text Key124829648
Report Number1213809-2018-00696
Device Sequence Number1
Product Code FMF
UDI-Device Identifier30382903096467
UDI-Public30382903096467
Combination Product (y/n)N
PMA/PMN Number
K980987
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type other,user facility
Type of Report Initial,Followup
Report Date 11/27/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Other
Device Expiration Date01/31/2023
Device Catalogue Number309646
Device Lot Number8035612
Initial Date Manufacturer Received 10/03/2018
Initial Date FDA Received10/23/2018
Supplement Dates Manufacturer Received10/03/2018
Supplement Dates FDA Received11/27/2018
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient Outcome(s) Other;
-
-