• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BARD ACCESS SYSTEMS SITE~RITE 6 SCANNER; SYSTEM, IMAGING, PULSED ECHO, ULTRASONIC

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BARD ACCESS SYSTEMS SITE~RITE 6 SCANNER; SYSTEM, IMAGING, PULSED ECHO, ULTRASONIC Back to Search Results
Catalog Number 9770066
Device Problem Image Display Error/Artifact (1304)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Type  malfunction  
Manufacturer Narrative
The manufacturer has received data that will be used for evaluation.Results are expected soon.A lot history review (lhr) review is not possible, as the device is manufactured using a unique serial number and not by lot number.A history review of serial number (b)(4) showed no other similar product complaint(s) from this serial number.
 
Event Description
Nurse reported system displays poor ultrasound image quality.
 
Manufacturer Narrative
The following were reviewed as part of this investigation: patient severity, applicable previous investigation(s), sample (if available), applicable fmea documents, labeling, and applicable manufacture records.Based on a review of this information, the following was concluded: the device was returned to the service facility for evaluation.During evaluation, the reported issue of poor image quality is confirmed.The root cause of the reported problem is burrs on the lvds cutout of the front metal case.The device was serviced, tested and returned to the customer.A history review of serial number dyyk7038 showed no other similar product complaint(s) from this serial number.
 
Event Description
Nurse reported system displays poor ultrasound image quality.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SITE~RITE 6 SCANNER
Type of Device
SYSTEM, IMAGING, PULSED ECHO, ULTRASONIC
Manufacturer (Section D)
BARD ACCESS SYSTEMS
605 n. 5600 w.
salt lake city UT 84116
MDR Report Key8001447
MDR Text Key124964787
Report Number3006260740-2018-03008
Device Sequence Number1
Product Code IYO
UDI-Device Identifier00801741091049
UDI-Public(01)00801741091049
Combination Product (y/n)N
PMA/PMN Number
K071204
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional,user faci
Type of Report Initial,Followup
Report Date 07/11/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received10/24/2018
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number9770066
Device Lot NumberN/A
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/15/2018
Event Location Hospital
Date Manufacturer Received06/17/2019
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-