Model Number 9732719
Device Problem Computer Software Problem (1112)
Patient Problem No Patient Involvement (2645)
Event Date 09/26/2018
Event Type  malfunction  
Manufacturer Narrative
A medtronic representative went to the site to test the equipment.Testing revealed that the software was displaying an error message randomly.This issue was resolved by reloading the software.The system then passed the system checkout and was found to be fully functional.Missing device manufacture date.If information is provided in the future, a supplemental report will be issued.
 
Event Description
Medtronic received information regarding an imaging device being used outside of procedure.It was reported that upon boot-up the site saw "an error has occurred that may damage system integrity".Site re-booted the system 3 times and the error still displayed, another re-boot resolved the issue.No patient present.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
N/A
Type of Device
IMAGE-INTENSIFIED FLUOROSCOPIC X-RAY SYSTEM, MOBILE
Manufacturer (Section D)
MEDTRONIC NAVIGATION, INC (LITTLETON)
300 foster st
littleton MA 01460
Manufacturer (Section G)
MEDTRONIC NAVIGATION, INC (LITTLETON)
300 foster st
littleton MA 01460
Manufacturer Contact
stacy ruemping
7000 central avenue ne rcw215
minneapolis, MN 55432
7635260594
MDR Report Key8005923
MDR Text Key125294594
Report Number3004785967-2018-00193
Device Sequence Number1
Product Code OXO
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K050996
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative
Reporter Occupation Other
Type of Report Initial
Report Date 10/25/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received10/25/2018
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number9732719
Device Catalogue Number9732719
Was Device Available for Evaluation? No
Date Manufacturer Received09/26/2018
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-