Catalog Number 295400-002 |
Device Problem
Electrical /Electronic Property Problem (1198)
|
Patient Problem
No Consequences Or Impact To Patient (2199)
|
Event Date 10/17/2018 |
Event Type
malfunction
|
Manufacturer Narrative
|
This alleged failure mode poses a low risk to the patient because the reported issue did not prevent the freedom driver from performing its life sustaining functions.In addition, the freedom driver has a redundant power source of multiple onboard batteries.The freedom hospital ac power supply will be returned to syncardia for evaluation.The results of the investigation will be provided in a follow-up mdr.(b)(4).
|
|
Event Description
|
The freedom ac power supply is a component that enables the freedom driver to be plugged into an external power source.The customer, a syncardia certified hospital, reported that the freedom hospital ac power supply did not connect properly into the freedom driver.
|
|
Manufacturer Narrative
|
The customer-reported issue, the inability to connect to a freedom driver, could not be reproduced or confirmed.Visual inspection did not identify any damage or abnormalities.The ac power supply passed all steps of functional inspection, including the ability to power a freedom driver for a minimum of one minute.This issue will continue to be monitored and trended as part of the customer experience process.Syncardia has completed its evaluation and is closing this file.Ce 4524 follow-up report 1.
|
|