• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ENCORE MEDICAL L.P. RSP SHOULDER; GLENOID, HEAD W/RETAINING SCREW, RSP, 36MM, NEUTRAL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ENCORE MEDICAL L.P. RSP SHOULDER; GLENOID, HEAD W/RETAINING SCREW, RSP, 36MM, NEUTRAL Back to Search Results
Model Number 508-36-101
Device Problem Break (1069)
Patient Problem Device Embedded In Tissue or Plaque (3165)
Event Date 10/03/2018
Event Type  Injury  
Event Description
Revision surgery - due to the central screw of glenoid baseplate breaking off into the glenoid.It was determined that the best course of action was to leave the well-fixed stem and convert prosthesis to a hemi-arthroplasty.
 
Manufacturer Narrative
The reason for this revision surgery was due to broken center screw of the baseplate.The previous revision surgery and the revision surgery detailed in this investigation occurred 6.9 months apart.There is no information in this complaint about any patient injuries, activities, or accidents that may have contributed to the need for this revision surgery.There are no reported pre-existing patient health conditions.The healthcare professional indicated there was no delay in surgery and another suitable device was available for use.The revision surgery was completed as intended.The device was disposed of at the hospital and not made available to djo surgical for examination.A review of the implant device history records (dhr), shows that the reported component used in the previous surgery, when released for use, met design and manufacturing requirements.There were no non-conforming material reports (ncmrs) associated with the product that may have contributed to the reported event.The device was verified to have gone through an acceptable sterilization process and was within its expiration date at the time of use during the previous surgery.Customer complaint history of the reported device showed no present trends or on-going issues that are in need of review.The root cause of this complaint was a revision surgery due to broken center screw of the baseplate.There were no findings during this investigation that indicate that the reported device was defective.There are multiple factors that may contribute to an event that are outside of the control of djo surgical.Inventory containment is not required as there are no indications of a product or process issue affecting implant safety or effectiveness.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
RSP SHOULDER
Type of Device
GLENOID, HEAD W/RETAINING SCREW, RSP, 36MM, NEUTRAL
Manufacturer (Section D)
ENCORE MEDICAL L.P.
9800 metric blvd
austin TX 78758 5445
MDR Report Key8007147
MDR Text Key125091210
Report Number1644408-2018-00983
Device Sequence Number1
Product Code PHX
UDI-Device Identifier00888912024761
UDI-Public(01)00888912024761
Combination Product (y/n)N
PMA/PMN Number
K051075
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative
Type of Report Initial,Followup
Report Date 12/20/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received10/25/2018
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Expiration Date01/30/2024
Device Model Number508-36-101
Device Catalogue Number508-36-101
Device Lot Number869C2131
Was Device Available for Evaluation? No
Date Manufacturer Received12/18/2018
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Treatment
506-03-114, LOT 831C1471; 506-03-114, LOT 831C1473; 506-03-126, LOT 834C1416; 506-03-134, LOT 836C1146; 508-32-104, LOT 866C2418
Patient Outcome(s) Required Intervention;
Patient Age63 YR
-
-