One used, list # 011-46110-71, transpac® iv monitoring kit w/03 ml squeeze flush device, microclave®, 03 ml safeset¿ reservoir and blood sampling port; lot # 3421156 was received for testing and evaluation.As received, the monitoring kit was hydrostatically pressure leak tested and a leak was observed at the bond between the 24" arterial tubing and female luer.The tip of the arterial tubing was microscopically examined and it appears that insufficient solvent was present on one half of the tubing, leading to a channel leak.The reported complaint of tubing detaching from a junction was confirmed.A channel leak was observed between arterial tubing and a luer of the used sample.It appears that that there was insufficient solvent at this bond location.The probable cause for the leak is a manual manufacturing process error during bonding.A device history review (dhr) was completed and no non-conformances were noted that would have contributed to the reported complaint.
|