Catalog Number 26-1230 |
Device Problem
Failure to Auto Stop (2938)
|
Patient Problem
Contusion (1787)
|
Event Date 10/10/2018 |
Event Type
Injury
|
Manufacturer Narrative
|
(b)(4).It has been reported that the device will be returned for evaluation.Upon completion of the investigation, a follow-up report will be submitted.
|
|
Event Description
|
As reported to codman via and (b)(6) report, the perforator drill bit did not disengage at the first trephine hole once the bone was pierced during a craniectomy.A new drill bit was utilized to continue the procedure.There was a contusion of the brain without gravity.There were no reports of delay.
|
|
Manufacturer Narrative
|
Additional information: udi (b)(4).Complaint sample was not returned to codman; therefore, an evaluation of the device could not be performed.A review of manufacturing records found no discrepancies related to the reported issue.The cause(s) of the difficulty reported by the customer could not be determined.If the complaint sample becomes available, this complaint will be reopened, and the respective evaluation performed.Trends will be monitored for this or similar complaints.At present, we consider this complaint to be closed.
|
|
Manufacturer Narrative
|
The device was received for evaluation.The perforator was visually inspected and no anomalies were observed.The perforator was then functionally tested.A series of holes with drilled without issues.The device performed as intended.A review of manufacturing records found no discrepancies when released to stock.Based on the investigation, the reported issue could not be confirmed.The device functioned as intended.Trends will be monitored for this and similar issues.At present, we consider this complaint to be closed.
|
|
Search Alerts/Recalls
|
|
|