• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: CODMAN & SHURTLEFF, INC. CODMAN CRANIOTOMY KIT; DRILLS, BURRS, TREPHINES & ACC.

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

CODMAN & SHURTLEFF, INC. CODMAN CRANIOTOMY KIT; DRILLS, BURRS, TREPHINES & ACC. Back to Search Results
Catalog Number 26-1230
Device Problem Failure to Auto Stop (2938)
Patient Problem Contusion (1787)
Event Date 10/10/2018
Event Type  Injury  
Manufacturer Narrative
(b)(4).It has been reported that the device will be returned for evaluation.Upon completion of the investigation, a follow-up report will be submitted.
 
Event Description
As reported to codman via and (b)(6) report, the perforator drill bit did not disengage at the first trephine hole once the bone was pierced during a craniectomy.A new drill bit was utilized to continue the procedure.There was a contusion of the brain without gravity.There were no reports of delay.
 
Manufacturer Narrative
Additional information: udi (b)(4).Complaint sample was not returned to codman; therefore, an evaluation of the device could not be performed.A review of manufacturing records found no discrepancies related to the reported issue.The cause(s) of the difficulty reported by the customer could not be determined.If the complaint sample becomes available, this complaint will be reopened, and the respective evaluation performed.Trends will be monitored for this or similar complaints.At present, we consider this complaint to be closed.
 
Manufacturer Narrative
The device was received for evaluation.The perforator was visually inspected and no anomalies were observed.The perforator was then functionally tested.A series of holes with drilled without issues.The device performed as intended.A review of manufacturing records found no discrepancies when released to stock.Based on the investigation, the reported issue could not be confirmed.The device functioned as intended.Trends will be monitored for this and similar issues.At present, we consider this complaint to be closed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CODMAN CRANIOTOMY KIT
Type of Device
DRILLS, BURRS, TREPHINES & ACC.
Manufacturer (Section D)
CODMAN & SHURTLEFF, INC.
325 paramount drive
raynham MA 02767
MDR Report Key8023745
MDR Text Key125640249
Report Number1226348-2018-10779
Device Sequence Number1
Product Code HBF
Combination Product (y/n)N
PMA/PMN Number
K933894
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type foreign,health professional
Type of Report Initial,Followup,Followup
Report Date 10/12/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received10/31/2018
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date05/31/2023
Device Catalogue Number26-1230
Device Lot NumberHY4195
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/13/2019
Date Manufacturer Received02/13/2019
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient Outcome(s) Other;
Patient Age37 YR
-
-