• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BOSTON SCIENTIFIC CORPORATION INTELLAMAP ORION HIGH RESOLUTION MAPPING CATHETER; CATHETER, ELECTRODE RECORDING, OR PROBE, ELECTRODE RECORDING

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC CORPORATION INTELLAMAP ORION HIGH RESOLUTION MAPPING CATHETER; CATHETER, ELECTRODE RECORDING, OR PROBE, ELECTRODE RECORDING Back to Search Results
Model Number M004RC64S0
Device Problems Signal Artifact/Noise (1036); Material Deformation (2976)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 08/17/2018
Event Type  malfunction  
Event Description
Reportable based on analysis completed on 02-oct-2018.It was reported that noisy proximal electrodes were noted during mapping.During a ventricular tachycardia ablation procedure with a intellamap orion catheter, more and more proximal electrode became noisy while mapping.The electrodes were disabled.The procedure was completed with another intellamap orion catheter.No patient complications were reported and the patient's current condition is fine.However, analysis revealed peeling of the deployment shaft.Visual inspection shows 3 kinked near the handle strain relief.The electrical test was performed and the device failed the test.Both the magnetic sensory resistance and inductance test are within specifications.The reported failure of proximal electrodes loss was confirmed through analysis.The device is seen to have organic damage cause by normal use of the device over a prolonged period.The deployment shaft peeling is present.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
INTELLAMAP ORION HIGH RESOLUTION MAPPING CATHETER
Type of Device
CATHETER, ELECTRODE RECORDING, OR PROBE, ELECTRODE RECORDING
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
4100 hamline avenue north
saint paul MN 55112
Manufacturer (Section G)
BOSTON SCIENTIFIC DE COSTA RICA S.R.L.
302 parkway
global park, la aurora,
heredia
CS  
Manufacturer Contact
timothy degroot
4100 hamline avenue north
dc a330,
saint paul, MN 55112
6515826168
MDR Report Key8025565
MDR Text Key125699566
Report Number2134265-2018-62326
Device Sequence Number1
Product Code DRF
UDI-Device Identifier08714729841968
UDI-Public08714729841968
Combination Product (y/n)N
Reporter Country CodeGM
PMA/PMN Number
K143481
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Physician
Type of Report Initial
Report Date 10/31/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date05/07/2019
Device Model NumberM004RC64S0
Device Catalogue Number87035
Device Lot Number0022086840
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer08/29/2018
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 10/02/2018
Initial Date FDA Received10/31/2018
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured05/07/2018
Is the Device Single Use? Yes
Type of Device Usage Initial
Patient Sequence Number1
-
-