ST. JUDE MEDICAL PUERTO RICO, INC. SJM REGENT HEART VALVE W/FLEX CUFF; HEART-VALVE, MECHANICAL Back to Search Results
Model Number 19AGFN-756
Device Problem Device Dislodged or Dislocated (2923)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 09/08/2018
Event Type  Injury  
Event Description
On (b)(6) 2018, a 19mm regent valve was selected.During device preparation, it was that noticed one of the leaflets had dislodged as a single piece from the butterfly hinges while rotating with the holder.The procedure was successfully completed with a 17mm regent valve as this was the next available size at the hospital.The procedure was extended by more than 30 minutes, but the patient was hemodynamically stable throughout the procedure and is reported stable.
 
Manufacturer Narrative
The reported event of a dislodged leaflet could not be confirmed.Both leaflets were properly inserted into the orifice and fully mobile.One chip was found on one recessed pivot area.It is unknown if the leaflet had been reinserted into the orifice.The device history record was reviewed to ensure that each manufacturing and inspection operation was performed.The cause of the reported event remains unknown; however, information from the field indicated that a smaller, 17mm, valve was subsequently implanted.Please note, per instructions for use artmt100122074 ver.A, " the valve should rotate freely.If resistance is noted, the valve holder/rotator may not be properly seated in the valve, the valve may not be in the fully closed position, or the valve may be oversized.If the valve does not rotate freely, do not force valve rotation.".
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SJM REGENT HEART VALVE W/FLEX CUFF
Type of Device
HEART-VALVE, MECHANICAL
Manufacturer (Section D)
ST. JUDE MEDICAL PUERTO RICO, INC.
p.o. box 998
lot 20 b st.
caguas, puerto rico 00725
MDR Report Key8029265
MDR Text Key125831019
Report Number2648612-2018-00089
Device Sequence Number1
Product Code LWQ
UDI-Device Identifier05414734005852
UDI-Public05414734005852
Combination Product (y/n)N
PMA/PMN Number
P810002
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Type of Report Initial,Followup
Report Date 01/17/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received11/01/2018
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Expiration Date05/31/2023
Device Model Number19AGFN-756
Device Catalogue Number19AGFN-756
Device Lot Number6481938
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/10/2018
Was the Report Sent to FDA? No
Date Manufacturer Received01/16/2019
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient Outcome(s) Hospitalization; Required Intervention;
Patient Age56 YR
Patient Weight67
-
-