Catalog Number 295400-001 |
Device Problem
Break (1069)
|
Patient Problem
No Consequences Or Impact To Patient (2199)
|
Event Date 10/23/2018 |
Event Type
malfunction
|
Manufacturer Narrative
|
This alleged failure mode poses a low risk to the patient because the reported issue did not prevent the freedom driver from performing its life sustaining functions.In addition, the freedom driver has a redundant power source of multiple onboard batteries.The freedom hospital ac power supply will be returned to syncardia for evaluation.The results of the investigation will be provided in a follow-up mdr.(b)(4).
|
|
Event Description
|
The freedom ac power supply is a component that enables the freedom driver to be plugged into an external power source.The customer, a syncardia certified hospital, reported that a piece of the freedom hospital ac power supply was broken.
|
|
Manufacturer Narrative
|
Visual inspection of the power supply revealed multiple chips, cracks and seam separation to the body of the power supply brick.The hypertronics connector was found to have internal and external damage resulting in the inability to keep the outer housing of the connector on the device.The root cause of this damage could not be conclusively determined, but given the observations made during the visual inspection, it is likely the result of a drop or other rough handling.This issue will continue to be monitored and trended as part of the customer experience process.Syncardia has completed its evaluation and is closing this file.(b)(4) follow-up report 1.
|
|