SYNCARDIA SYSTEMS, LLC SYNCARDIA FREEDOM HOSPITAL AC POWER SUPPLY Back to Search Results
Catalog Number 295400-001
Device Problem Break (1069)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 10/23/2018
Event Type  malfunction  
Manufacturer Narrative
This alleged failure mode poses a low risk to the patient because the reported issue did not prevent the freedom driver from performing its life sustaining functions.In addition, the freedom driver has a redundant power source of multiple onboard batteries.The freedom hospital ac power supply will be returned to syncardia for evaluation.The results of the investigation will be provided in a follow-up mdr.(b)(4).
 
Event Description
The freedom ac power supply is a component that enables the freedom driver to be plugged into an external power source.The customer, a syncardia certified hospital, reported that a piece of the freedom hospital ac power supply was broken.
 
Manufacturer Narrative
Visual inspection of the power supply revealed multiple chips, cracks and seam separation to the body of the power supply brick.The hypertronics connector was found to have internal and external damage resulting in the inability to keep the outer housing of the connector on the device.The root cause of this damage could not be conclusively determined, but given the observations made during the visual inspection, it is likely the result of a drop or other rough handling.This issue will continue to be monitored and trended as part of the customer experience process.Syncardia has completed its evaluation and is closing this file.(b)(4) follow-up report 1.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SYNCARDIA FREEDOM HOSPITAL AC POWER SUPPLY
Type of Device
AC POWER SUPPLY
Manufacturer (Section D)
SYNCARDIA SYSTEMS, LLC
1992 e. silverlake road
tucson AZ 85713
MDR Report Key8033902
MDR Text Key127886893
Report Number3003761017-2018-00523
Device Sequence Number1
Product Code LOZ
Combination Product (y/n)N
PMA/PMN Number
P030011
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional,user faci
Type of Report Initial,Followup
Report Date 01/16/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number295400-001
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/09/2018
Initial Date Manufacturer Received 10/23/2018
Initial Date FDA Received11/02/2018
Supplement Dates Manufacturer Received10/23/2018
Supplement Dates FDA Received01/17/2019
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient Age54 YR
-
-