• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: WRIGHTS LANE SYNTHES USA PRODUCTS LLC DEPTH GAUGE FOR 2.0MM AND 2.4MM SCREWS; GAUGE,DEPTH

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

WRIGHTS LANE SYNTHES USA PRODUCTS LLC DEPTH GAUGE FOR 2.0MM AND 2.4MM SCREWS; GAUGE,DEPTH Back to Search Results
Catalog Number 319.006
Device Problem Break (1069)
Patient Problem No Patient Involvement (2645)
Event Type  malfunction  
Manufacturer Narrative
Complainant part is expected to be returned for manufacturer review/investigation, but has yet to be received.Reporter is synthes sales consultant.A review of the device history records has been requested.The investigation could not be completed; no conclusion could be drawn, as no product was received.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event Description
It was reported that during an unknown event on an unknown date, the tip of the depth gauge for 2.0mm and 2.4mm screws has broken off.There was no patient involvement.This report is for one (1) depth gauge for 2.0mm and 2.4mm screws.This is report 1 of 1 for complaint (b)(4).
 
Manufacturer Narrative
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.Device history records review was completed for part: 319.006, lot: 5069138.Manufacturing location: brandywine, release to warehouse date: aug 20, 2005.No non conformance reports were generated during production.Review of the device history records showed that there were no issues during the manufacture of the product that would contribute to this complaint condition.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Manufacturer Narrative
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.Device history lot, dhr review for part # 319.006 , synthes lot # 5069138 , release to warehouse date 20-aug-2005 , manufactured by synthes brandywine.The material was reviewed and the hardness value was confirmed to meet the specification with no (relevant) non-conformance noted.No ncrs were generated during production.Review of the device history record(s) showed that there were no issues during the manufacture of this product that would contribute to this complaint condition.Product was not returned.Based on the information available, it has been determined that no corrective and preventative action is proposed.This complaint will be accounted for and monitored via post market surveillance activities.If additional information is made available, the investigation will be updated as applicable.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
DEPTH GAUGE FOR 2.0MM AND 2.4MM SCREWS
Type of Device
GAUGE,DEPTH
Manufacturer (Section D)
WRIGHTS LANE SYNTHES USA PRODUCTS LLC
1302 wrights lane east
west chester PA 19380
MDR Report Key8043618
MDR Text Key127878581
Report Number2939274-2018-54869
Device Sequence Number1
Product Code HTJ
UDI-Device Identifier10886982189943
UDI-Public(01)10886982189943
Combination Product (y/n)N
PMA/PMN Number
EXEMPT
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative
Type of Report Initial,Followup,Followup
Report Date 10/23/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number319.006
Device Lot Number5069138
Was Device Available for Evaluation? No
Initial Date Manufacturer Received 10/23/2018
Initial Date FDA Received11/06/2018
Supplement Dates Manufacturer Received11/12/2018
12/12/2018
Supplement Dates FDA Received11/29/2018
12/12/2018
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-