Model Number 3100 A
Device Problem Overheating of Device (1437)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 10/06/2018
Event Type  Injury  
Manufacturer Narrative
Results of investigation: this unit was field serviced by a vyaire medical field service technician.The service technician was able to verify the customer's reported issue during testing and evaluation.A component on the alarm pcb was discolored and appeared to have solder leaking from the inside.The service technician turned on the unit and was able to smell the smoke from the head of the unit.The service technician will change the pcb board to address the customer's reported issue.
 
Event Description
It was reported to vyaire medical that the unit overheated and the stopped while in use on a patient.The unit alarmed and there was an electrical/burning smell.The patient was placed on another unit with no report of any patient harm.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
3100 HIGH FREQUENCY OSCILLATING VENTILATOR (HFOV)
Type of Device
VENTILATOR, CONTINUOUS, FACILITY USE
Manufacturer (Section D)
VYAIRE MEDICAL
26125 north riverwoods blvd
mettawa IL 60045
Manufacturer (Section G)
VYAIRE MEDICAL INC.
1100 bird center drive
palm springs CA 92262
Manufacturer Contact
mindy faber
26125 north riverwoods blvd
mettawa, IL 60045
8727570116
MDR Report Key8044188
MDR Text Key126348053
Report Number2021710-2018-09526
Device Sequence Number1
Product Code LSZ
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P890057
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional,user faci
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 11/06/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received11/06/2018
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number3100 A
Device Catalogue Number768901
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Date Manufacturer Received10/09/2018
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured07/01/2002
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
-
-