The reason for this revision surgery was due to dislocation.The previous surgery and the revision detailed in this investigation occurred 6 months apart.There is no information in this complaint about any patient injuries, activities, or accidents that may have contributed to the need for this revision surgery.There are no reported pre-existing patient health conditions.The healthcare professional indicated there was no delay in surgery and another suitable device was available for use.The revision surgery was completed as intended.The device was disposed of at the hospital and not made available to djo surgical for examination.A review of the implant device history records (dhr) shows that the reported component used in the previous surgery met design and manufacturing requirements.There was a non-conforming material report (ncmr) # 41215 associated with the part# 508-36-101, glenoid, head with retaining screw, reverse shoulder prosthesis (rsp) 36millimeter,neutral, which documents a nonconformance that out of 15 quantity lot 1 item was rejected, due to handling, it was reworked with suitable operations and accepted.All items in the lot were met design, fit and function requirements and were accepted.Customer complaint history of the reported device showed no present trends or on-going issues that are in need of review.He root cause of this complaint was a revision surgery due to dislocation.There were no findings during this investigation that indicate that the reported device was the source or had a direct connection with the event.There are many factors that may contribute to the event that are outside the control of djo surgical such as: patient activities, inadequate soft tissue support, excessive range of motion, patient bone deterioration or trauma.Inventory containment is not required as there are no indications of a product or process issue affecting implant safety or effectiveness.
|