• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ZIMMER MANUFACTURING B.V. LINER 26 MM I.D. FOR USE WITH 42/43 MM O.D. SHELLS; PROSTHESIS, HIP

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ZIMMER MANUFACTURING B.V. LINER 26 MM I.D. FOR USE WITH 42/43 MM O.D. SHELLS; PROSTHESIS, HIP Back to Search Results
Model Number N/A
Device Problem Difficult to Insert (1316)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 05/23/2018
Event Type  malfunction  
Manufacturer Narrative
(b)(4).Foreign source: (b)(6).Complaint sample was evaluated and the reported event was confirmed.Visual inspection determined that the poly ring was deformed and no longer in its intended position.Dimensional analysis could not be performed as damage was too severe.No other anomalies were noted.Device history record (dhr) was reviewed and no discrepancies relevant to the reported event were found.Root cause was unable to be determined as the necessary information to adequately investigate the reported event was not provided.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Event Description
It was reported that during total hip arthroplasty, the locking ring of the multipolar liner deviated from the groove.This surgery was finished with backup product.No adverse event noted to patient.Attempts have been made to obtain additional information; however, none is available.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LINER 26 MM I.D. FOR USE WITH 42/43 MM O.D. SHELLS
Type of Device
PROSTHESIS, HIP
Manufacturer (Section D)
ZIMMER MANUFACTURING B.V.
turpeaux industrial park
route #1 km 123.4 bldg #1
mercedita PR 00715
Manufacturer (Section G)
ZIMMER MANUFACTURING B.V.
turpeaux industrial park
route #1 km 123.4 bldg #1
mercedita PR 00715
Manufacturer Contact
christina arnt
56 e. bell dr.
warsaw, IN 46582
5745273773
MDR Report Key8049295
MDR Text Key126515204
Report Number0002648920-2018-00804
Device Sequence Number1
Product Code KWY
Combination Product (y/n)N
Reporter Country CodeJA
PMA/PMN Number
K833991
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Physician
Type of Report Initial
Report Date 11/07/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date03/31/2024
Device Model NumberN/A
Device Catalogue Number00500104226
Device Lot Number63295549
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer07/24/2018
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 06/01/2018
Initial Date FDA Received11/07/2018
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured04/06/2016
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage N
Patient Sequence Number1
-
-