It was reported that patient's device showed high impedance.Patient underwent lead replacement surgery.During surgery, lead pin was removed, cleaned and re-inserted but high impedance did not resolve.Lead pin insertion was checked but the high impedance did not resolve until lead was replaced.After lead replacement.The explanted lead was received.Analysis is underway but has not been completed to date.
|
An abraded opening was noted in the outer silicone tubing.A continuity check performed between the connector pin and the connector ring identified a short circuit condition as a result of the lead coils protruding out of the inner tubing.Based on the appearance of the returned lead portion this condition was most likely caused at explant.The reported lead fracture was not verified.Based on the appearance of the returned lead portion, it is believed that the identified kinks and outer tubing cut/tear openings, stretched coils, and coils protruding out from the inner and the outer silicone tubing, and contact between the positive and the negative lead coils were most likely caused during the explant procedure.Since the electrode array portion was not returned for analysis, an evaluation and resulting commentary cannot be made on that portion of the lead.Other than the above mentioned observations and typical wear and explant related observations, no other anomalies were identified in the returned lead portion.
|