• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: HILL-ROM BATESVILLE ADVANTA FRAME; BED, AC-POWERED ADJUSTABLE HOSPITAL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

HILL-ROM BATESVILLE ADVANTA FRAME; BED, AC-POWERED ADJUSTABLE HOSPITAL Back to Search Results
Model Number P1600B005348
Device Problem Device Alarm System (1012)
Patient Problem No Patient Involvement (2645)
Event Date 10/29/2018
Event Type  malfunction  
Manufacturer Narrative
The account found the bed exit siderail interface and the communication cable needed to be replaced.Per the hill-rom service manual the advanta bed requires an effective maintenance program.Preventative maintenance will minimize downtime due to excessive wear.Communications: inspect and test the communication junction box.Test the sidecom communication system features for proper operation.Inspect the communication cable including the male and female pins in the plug.Test the nurse call super capacitor for proper operation.A search of the hill-rom maintenance records did not show hill-rom performed any preventative maintenance on this bed.It is unknown if the facility performs preventative maintenance on their beds.The account replaced the bed exit siderail interface and the communication cable to resolve the issue.Based on this information, no further action is required.
 
Event Description
Hill-rom technical support received a report from the account stating the bed exit alarm was not alarming at the nurses' station.The bed was located in the med surg department at the account.There was no patient/user injury reported.This report was filed in our complaint handling system as complaint #(b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ADVANTA FRAME
Type of Device
BED, AC-POWERED ADJUSTABLE HOSPITAL
Manufacturer (Section D)
HILL-ROM BATESVILLE
1069 state route 46 east
batesville IN 47006
Manufacturer Contact
lori daulton
1069 state route 46 east
batesville, IN 47006
8129312280
MDR Report Key8059635
MDR Text Key127870766
Report Number1824206-2018-00434
Device Sequence Number1
Product Code FNL
Combination Product (y/n)N
Reporter Country CodeUS
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative
Reporter Occupation Other
Type of Report Initial
Report Date 10/29/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received11/12/2018
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator No Information
Device Model NumberP1600B005348
Was Device Available for Evaluation? No
Was the Report Sent to FDA? No
Date Manufacturer Received10/29/2018
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-