Additional information: upon receipt, interrogation of the device revealed the device was above eri when received.The device was tested on the bench using clinical leads, test leads, resistor loads, and pin gauges.Both test leads and clinical leads were fully inserted and secured without difficulty.Telemetry, pli, hvli were tested and no anomalies were observed.All lead bore gauge measurements were within specifications.No anomaly was detected.The reported field event of lead insertion anomaly was not confirmed.
|