Catalog Number 180735-1
Device Problem Adverse Event Without Identified Device or Use Problem (2993)
Patient Problems Pain (1994); Injury (2348)
Event Date 10/17/2018
Event Type  Injury  
Manufacturer Narrative
Review of the device history records indicate devices were manufactured and accepted into final stock with no reported discrepancies.There have been no other similar events for the lot referenced.The following devices were also listed in this report: mck femoral-rm-ll-sz 4, cat# 180514, lot# 231345k-1.Mck tibial baseplate-rm/ll-sz 5, cat# 180615, lot# 26240516-01.It cannot be determined which, if any of these devices may have caused or contributed to the patient's experience.An evaluation of the device cannot be performed as the device was not returned to the manufacturer.Should additional information become available it will be reported in a supplemental report upon completion of the investigation.
 
Event Description
Dr.(b)(6) revised a mako uni knee to a primary cementless tka due to pain.(b)(6) observed the patellafemoral joint disease progressed a bit.Implants were not loose.
 
Manufacturer Narrative
An event regarding revision due to patient factors (disease progression) involving a mako insert was reported.The event was confirmed through clinician review of the medical records provided.Method & results: product evaluation and results: not performed as no product was returned for evaluation.Clinician review: advanced degenerative disease in the patellofemoral compartment of a knee with a medial unicompartmental replacement is confirmed.Insufficient information has been received to confirm that this was progressive.No details of the revision surgery were provided.Product history review: indicated all devices were manufactured and accepted into final stock with no relevant reported discrepancies.Complaint history review: there have been no other similar events for the lot referenced.Conclusion: advanced degenerative disease in the patellofemoral compartment of a knee with a medial unicompartmental replacement is confirmed.Insufficient information has been received to confirm that this was progressive.No details of the revision surgery were provided.Further information such as operative notes, pre and post operative x-rays, imaging from the index and revision surgeries as well as outpatient office and clinic notes are required to complete the investigation for determining a root cause.No further investigation is required at this time.If additional information becomes available to indicate further evaluation is warranted, this record will be reopened.
 
Event Description
Dr.(b)(6) revised a mako uni knee to a primary cementless tka due to pain.Marchand observed the patellafemoral joint disease progressed a bit.Implants were not loose.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MAKO X3 UNI ONLAY TIBIAL INSERT SIZE 5 - 8 MM
Type of Device
PROSTHESIS, KNEE PATELLOFEMOROTIBIAL, PARTIAL, SEMI-CONSTRAINED, CEMENTED, POLYM
Manufacturer (Section D)
MAKO SURGICAL CORP.
2555 davie road
fort lauderdale FL 33317
MDR Report Key8065747
MDR Text Key127059206
Report Number3005985723-2018-00673
Device Sequence Number1
Product Code NPJ
UDI-Device Identifier00848486016548
UDI-Public00848486016548
Combination Product (y/n)N
PMA/PMN Number
K150307
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional,other
Type of Report Initial,Followup
Report Date 01/07/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Expiration Date12/31/2021
Device Catalogue Number180735-1
Device Lot NumberM063N8
Was Device Available for Evaluation? No
Initial Date Manufacturer Received 10/17/2018
Initial Date FDA Received11/13/2018
Supplement Dates Manufacturer Received12/10/2018
Supplement Dates FDA Received01/07/2019
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient Outcome(s) Hospitalization; Required Intervention;
Patient Age68 YR
-
-