MALEM MEDICAL LTD MALEM ; "ALARM, CONDITIONED RESPONSE ENURESIS" Back to Search Results
Model Number M04B
Device Problems Leak/Splash (1354); Electrical Shorting (2926)
Patient Problems Bruise/Contusion (1754); Scarring (2061); Burn, Thermal (2530)
Event Date 11/05/2018
Event Type  Injury  
Event Description
Our clinic treated a (b)(6) boy who has been burnt from an enuresis alarm.The child is stable but has big bruises on his neck where the alarm was touching his skin.This is the result of an electrical malfunction in the alarm.The device short circuited and caused batteries to leak.The batteries spilled onto the child's body.A combination of hot plastic and battery scarred the child's skin while he was asleep.We have recommended that the parents discontinue using the alarm.We have returned the alarm back to the parents.The alarm housing has bent inwards from excess heat.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MALEM
Type of Device
"ALARM, CONDITIONED RESPONSE ENURESIS"
Manufacturer (Section D)
MALEM MEDICAL LTD
lowdham, nottingham NG147 EJ
UK  NG147EJ
MDR Report Key8074490
MDR Text Key127532132
Report NumberMW5081297
Device Sequence Number1
Product Code KPN
Combination Product (y/n)N
Reporter Country CodeCA
Number of Events Reported1
Summary Report (Y/N)N
Report Source Voluntary
Reporter Occupation Pharmacist
Type of Report Initial
Report Date 11/09/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Device Operator Other
Device Model NumberM04B
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received Not provided
Initial Date FDA Received11/14/2018
Was Device Evaluated by Manufacturer? No Information
Type of Device Usage N
Patient Sequence Number1
Patient Outcome(s) Hospitalization;
Patient Age4 YR
Patient Weight17
-
-