Model Number 6101
Device Problem Device Displays Incorrect Message (2591)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Type  malfunction  
Event Description
Information was received that the cadd prizm vip pump had error 10102.No reported adverse effects.
 
Manufacturer Narrative
One cadd prizm vip pump was returned for analysis.Visual inspection performed, and product found to be in a good condition.Event history log review was performed and found no evidence of reported problem.During investigation the pump was visually inspected the and battery inserted onto pump during power up and its alarm "error code 10102".Investigator was also unable performed any other test.The customer's stated problem was verified.Inspected the pump and power up the device and it's alarmed with error code 10102.Further investigation of the pump and determined that a faulty microprocessor board was the root cause of the issue.Service replaced the faulty microprocessor board to correct the reported problem.The root cause of the reported problem was the faulty microprocessor board.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CADD-PRIZM VIP SYSTEM
Type of Device
PUMP, INFUSION, PCA
Manufacturer (Section D)
SMITHS MEDICAL ASD, INC.
6000 nathan lane north
minneapolis MN 55442
Manufacturer (Section G)
SMITHS MEDICAL ASD, INC.
1265 grey fox road
st. paul MN 55112
Manufacturer Contact
dave halverson
6000 nathan lane north
minneapolis, MN 55442
7633833310
MDR Report Key8076345
MDR Text Key127358310
Report Number3012307300-2018-08313
Device Sequence Number1
Product Code MEA
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K000842
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type distributor
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 01/03/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received11/15/2018
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator No Information
Device Model Number6101
Device Catalogue Number21-8821-01
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/22/2018
Is the Reporter a Health Professional? No
Date Manufacturer Received12/04/2018
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
-
-