• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: COVIDIEN ADULT MULTI-FUNCTION DEFIBRILLATION ELECTRODES; PACEMAKER, CARDIAC, EXTERNAL, TRANSCUTANEOUS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

COVIDIEN ADULT MULTI-FUNCTION DEFIBRILLATION ELECTRODES; PACEMAKER, CARDIAC, EXTERNAL, TRANSCUTANEOUS Back to Search Results
Catalog Number 22550A
Device Problems Intermittent Continuity (1121); Device Displays Incorrect Message (2591); Device Sensing Problem (2917)
Patient Problem No Code Available (3191)
Event Date 11/14/2018
Event Type  Injury  
Event Description
Defib patches are not functioning correctly.When first applied the patches are recognized by the defibrillator and a rhythm is displayed.Intermittently the rhythm is lost and the defibrillator says "connect electrodes".This has happened on multiple defibrillators but seems to happen with the specific "covidien medi-trace cadence" brand of patches.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ADULT MULTI-FUNCTION DEFIBRILLATION ELECTRODES
Type of Device
PACEMAKER, CARDIAC, EXTERNAL, TRANSCUTANEOUS
Manufacturer (Section D)
COVIDIEN
MDR Report Key8077989
MDR Text Key127892258
Report NumberMW5081356
Device Sequence Number1
Product Code DRO
Combination Product (y/n)N
Reporter Country CodeUS
Number of Events Reported1
Summary Report (Y/N)N
Report Source Voluntary
Reporter Occupation Nurse
Type of Report Initial
Report Date 11/14/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date07/23/2020
Device Catalogue Number22550A
Device Lot Number821838X
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received Not provided
Initial Date FDA Received11/15/2018
Was Device Evaluated by Manufacturer? No Information
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage N
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
-
-