• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SMITHS MEDICAL ASD, INC. CADD-PRIZM VIP SYSTEM; PUMP, INFUSION, PCA

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SMITHS MEDICAL ASD, INC. CADD-PRIZM VIP SYSTEM; PUMP, INFUSION, PCA Back to Search Results
Model Number 6101
Device Problem Device Displays Incorrect Message (2591)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Type  malfunction  
Event Description
Information was received that the cadd prizm vip pump had constant wrong cassette alarms.No reported adverse effects.
 
Manufacturer Narrative
One cadd prizm vip pump was returned for analysis.Visual inspection performed and product found to be in good condition.Customer's complaint of wrong cassette, will not run was verified.The event log showed wrong cassette alarms.Service replaced the downstream occlusion sensor.The problem source is user unduced fluid ingression with the downstream occlusion sensor.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CADD-PRIZM VIP SYSTEM
Type of Device
PUMP, INFUSION, PCA
Manufacturer (Section D)
SMITHS MEDICAL ASD, INC.
6000 nathan lane north
minneapolis MN 55442
MDR Report Key8079528
MDR Text Key127639157
Report Number3012307300-2018-08310
Device Sequence Number1
Product Code MEA
Combination Product (y/n)N
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type distributor
Type of Report Initial,Followup
Report Date 12/28/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received11/16/2018
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator No Information
Device Model Number6101
Device Catalogue Number21-8821-01
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/22/2018
Date Manufacturer Received11/28/2018
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-