• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: C.R. BARD, INC. (COVINGTON) -1018233 BULB TYPE IRRIGATION SYRINGE 50ML, PLASTIC STERILE; BULB SYRINGE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

C.R. BARD, INC. (COVINGTON) -1018233 BULB TYPE IRRIGATION SYRINGE 50ML, PLASTIC STERILE; BULB SYRINGE Back to Search Results
Catalog Number 0035280
Device Problems Device Contamination with Chemical or Other Material (2944); Appropriate Term/Code Not Available (3191)
Patient Problems No Consequences Or Impact To Patient (2199); No Known Impact Or Consequence To Patient (2692)
Event Date 10/22/2018
Event Type  malfunction  
Manufacturer Narrative
The investigation is still in progress.Once the investigation is complete a supplemental report will be filed.
 
Event Description
It was reported that green foreign material came out of the inside of the syringe during use.
 
Event Description
It was reported that green foreign material came out of the inside of the syringe during use.
 
Manufacturer Narrative
The reported event was confirmed.Visual evaluation of the returned sample noted one opened bulb irrigation syringes in the original packaging.A green silicone substance (0.7615in long) found outside the syringe.The device history record was reviewed and found nothing that could have caused or contributed to the reported event.The instructions for use were found adequate and state the following: ¿this is a single use device.Do not re-sterilize any portion of this device.Reuse and/or repackaging may create a risk of patient or user infection, compromise the structural integrity and/or essential material and design characteristics of the device, which may lead to device failure, and/or lead to injury, illness or death of the patient.Contains or presence of phthalates: di(2-ethylhexyl)phthalate (dehp) is a plasticizer used in some polyvinyl chloride medical devices.Dehp has been shown to produce a range of adverse effects in experimental animals, notably liver toxicity and testicular atrophy.Although the toxic and carcinogenic effects of dehp have been well established in experimental animals, the ability of this compound to produce adverse effects in humans is controversial.There is no evidence that neonates, infants, pregnant and breast feeding women exposed to dehp experience any related adverse effects.However, a lack of evidence of causation between dehp-pvc and any disease or adverse effect does not mean that there are no risks.".
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BULB TYPE IRRIGATION SYRINGE 50ML, PLASTIC STERILE
Type of Device
BULB SYRINGE
Manufacturer (Section D)
C.R. BARD, INC. (COVINGTON) -1018233
8195 industrial blvd
covington GA 30014
MDR Report Key8083117
MDR Text Key127560917
Report Number1018233-2018-05510
Device Sequence Number1
Product Code KYZ
UDI-Device Identifier00801741049255
UDI-Public(01)00801741049255
Combination Product (y/n)N
PMA/PMN Number
EXEMPT
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type distributor,other
Type of Report Initial,Followup
Report Date 05/14/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received11/18/2018
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date05/31/2023
Device Catalogue Number0035280
Device Lot NumberNGCS2581
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/10/2018
Date Manufacturer Received05/01/2019
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-