• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DEPUY INTERNATIONAL LTD - 8010379 ASR UNI FEMORAL IMPL SIZE 51; ASR HIP SYSTEM : HIP METAL FEMORAL HEADS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEPUY INTERNATIONAL LTD - 8010379 ASR UNI FEMORAL IMPL SIZE 51; ASR HIP SYSTEM : HIP METAL FEMORAL HEADS Back to Search Results
Catalog Number 999890251
Device Problems Corroded (1131); Degraded (1153); Naturally Worn (2988); Noise, Audible (3273)
Patient Problems Host-Tissue Reaction (1297); Edema (1820); Foreign Body Reaction (1868); Hypersensitivity/Allergic reaction (1907); Necrosis (1971); Pain (1994); Pocket Erosion (2013); Swelling (2091); Tissue Damage (2104); Discomfort (2330); Injury (2348); Joint Disorder (2373); Ambulation Difficulties (2544); Limited Mobility Of The Implanted Joint (2671); Test Result (2695); No Code Available (3191)
Event Date 07/17/2017
Event Type  Injury  
Manufacturer Narrative
Product complaint # (b)(4).If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.No pma/510(k) number provided because this implant is sold internationally with different indications for use; it is currently sold in the us under a different part number.The correction/removal reporting number listed applies to the corresponding product code sold domestically.Concomitant medical products: asr uni femoral impl size 51 hip asr femoral head; asr acetabular cups 58 hip asr acetabular cup; adapter sleeves 12/14 +5 hip asr femoral sleeve/augment; summit duofix tap sz4 hi off hip femoral stem.Initial reporter occupation: non-healthcare professional ¿ attorney.(b)(4).
 
Event Description
New unity record created in order to update etq complaint number (b)(4).Patient was revised to address removal of the asr for unknown reason.It also states that patient had other side revised.It was mom liner.Update ad 25 oct 2018: (b)(4) is a re-opened record of (b)(4) due to receipt of litigation records.Litigation alleged increasing pain, swelling, elevated levels of cobalt and chromium in the patient's bloodstream, injury, discomfort, poor balance, difficulty walking, popping, clicking and grinding noises, bone erosion, femoral acetabular loosening, impingement and/or detachment, metallosis, emotional distress and suffering.Added patient harms, added stem as per new allegation of "elevated levels of cobalt and chromium in the patient's bloodstream." and affected hip.Doi: (b)(6) 2009; dor: (b)(6) 2017; right hip.
 
Manufacturer Narrative
Product complaint # (b)(4).Investigation summary: the asr platform was voluntarily recalled from the market in august 2010, and the asr product codes are now considered inactive.Further investigation of this individual incident will not be undertaken.The investigation regarding the root cause(s) and/or corrective actions was conducted under mdd capa(b)(4).Ongoing post market surveillance is conducted per our procedures for this product if information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Manufacturer Narrative
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.H10 additional narrative: added: b1, h6 (patient).Corrected: h6 (device).H6 patient code: no code available (3191) was used to capture emotional distress.Product complaint # (b)(4).Investigation summary: no device associated with this report was received for examination.This hip replacement platform was voluntarily recalled from the market and the product codes are now considered inactive.Depuy considers the investigation closed.Should additional information be received, the information will be reviewed and the investigation will be re-opened as necessary.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Manufacturer Narrative
Product complaint # (b)(4).Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees caused or contributed to the potential event described in this report.No code available use for device revision or replacement.
 
Event Description
After review of medical records patient was revised to addressed failed right total hip arthroplasty with trunnionosis and abductor necrosis.Operative notes indicated consistent pseudotumor formation, fluid stained and extensive evidence of metal debris, minimal bone loss and evidence of corrosion at the taper and bearing surface of femoral head.Added expiration date, medical history and date of birth of patient.Corrected patient's initial.Doi: (b)(6) 2009 dor: (b)(6) 2017 right hip.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ASR UNI FEMORAL IMPL SIZE 51
Type of Device
ASR HIP SYSTEM : HIP METAL FEMORAL HEADS
Manufacturer (Section D)
DEPUY INTERNATIONAL LTD - 8010379
st. anthony's road
leeds LS11 8 DT
UK  LS11 8 DT
MDR Report Key8083958
MDR Text Key127543463
Report Number1818910-2018-75870
Device Sequence Number1
Product Code KXA
Combination Product (y/n)N
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type other
Remedial Action Recall
Type of Report Initial,Followup,Followup,Followup
Report Date 10/25/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date03/31/2019
Device Catalogue Number999890251
Device Lot Number2864993
Was Device Available for Evaluation? No
Initial Date Manufacturer Received 10/25/2018
Initial Date FDA Received11/19/2018
Supplement Dates Manufacturer Received12/11/2018
04/29/2020
05/20/2020
Supplement Dates FDA Received12/13/2018
05/15/2020
05/21/2020
Is This a Reprocessed and Reused Single-Use Device? No
Removal/Correction NumberZ-1749/1816-2011
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
Patient Age68 YR
-
-