Catalog Number 6487800
Device Problem Loss of Power (1475)
Patient Problem No Patient Involvement (2645)
Event Type  malfunction  
Manufacturer Narrative
(b)(4).
 
Event Description
It was reported that the ventilator shut down prematurely during transport.There was no patient involvement.(b)(4).
 
Manufacturer Narrative
The biomed at the user facility replaced the external batteries.Our fse (field service engineer) then was on site and investigated the reported issue.The fse exchanged the two internal batteries then the ventilator passed all tests and was cleared for clinical use.The reported shut down and preceding battery related alarms can be confirmed in the logs due to expired internal battery.The replaced batteries were not returned for our investigation.(b)(4).
 
Event Description
(b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SERVO-I
Type of Device
VENTILATOR, CONTINUOUS, FACILITY USE
Manufacturer (Section D)
MAQUET CRITICAL CARE AB
solna
SW 
Manufacturer (Section G)
MAGNUS LINDQVIST
maquet critical care ab
röntgenvägen 2, se-171 54
solna
SW  
Manufacturer Contact
maquet critical care ab
röntgenvägen 2, se-171 54
solna 
MDR Report Key8090264
MDR Text Key128198346
Report Number8010042-2018-00613
Device Sequence Number1
Product Code CBK
Combination Product (y/n)N
PMA/PMN Number
K123149
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Biomedical Engineer
Type of Report Initial,Followup
Report Date 12/21/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received11/20/2018
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number6487800
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received10/30/2018
Was Device Evaluated by Manufacturer? No
Date Device Manufactured08/06/2011
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage N