• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: VYAIRE MEDICAL 3100A VENTILATOR; VENTILATOR, HIGH FREQUENCY

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

VYAIRE MEDICAL 3100A VENTILATOR; VENTILATOR, HIGH FREQUENCY Back to Search Results
Model Number 3100A
Device Problem Failure to Cycle (1142)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 10/29/2018
Event Type  Injury  
Manufacturer Narrative
Field service rep device evaluation: the field service engineer (fse) went on-site to evaluate the suspect device.The fse cycled the device on and could not duplicate the reported device behavior.The fse performed the patient circuit calibration and performance verification, which passed.Having met manufacture specifications, the device was return for use.At this time, no assembly failure was identified therefore, no component root cause investigation can be performed.
 
Event Description
The customer reported this ventilator device stopped oscillating, while in patient use.The customer stated no patient harm or injury with this event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
3100A VENTILATOR
Type of Device
VENTILATOR, HIGH FREQUENCY
Manufacturer (Section D)
VYAIRE MEDICAL
22745 savi ranch pkwy
yorba linda CA 92887
Manufacturer (Section G)
VYAIRE MEDICAL INC.
1100 bird center drive
palm springs CA 92262
Manufacturer Contact
stanley tan
22745 savi ranch pkwy
yorba linda, CA 92887
7149193324
MDR Report Key8090650
MDR Text Key127925350
Report Number2021710-2018-09582
Device Sequence Number1
Product Code LSZ
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P890057
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional,user faci
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 10/29/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number3100A
Device Catalogue Number768901
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 10/29/2018
Initial Date FDA Received11/20/2018
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured10/01/2002
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
-
-